Hi All
there is no intention now nor has there ever been one to separate 3D shimming and charging for it. Topshim 3d is an integral part of our software.
It has been around since day one when it was called gradshim3d.
For those who do not want to spend 20 minutes you can get the quick version by using the "3dfast" option. This will do not more than 3 iterations instead of running to convergence. One an already well shimmed system this will take ca 5 minutes.
Best regards
Clemens
On 9/25/2018 2:46 PM, CHARLES G FRY wrote:
Hi Bobby,
This is curious. 3D gradient shimming has always in the past been available at no charge in TopSpin. It is definitely a useful feature at any point where the probe has been worked on, and when it is being first installed. It is easy to use, and does a good job. We use it regularly to help us update our saved shims, something we do about monthly. It is from those saved shims that we then do the 1D gradient shimming on the in-between. The key here is that it pulls in good values for the non-axial (non-Z) shims. These can be very hard to manually shim in. Back in the old days, it would take 1-2 days to shim in a new probe. Now it takes maybe an hour. Updating a shim set used to take me an hour or more (especially if the probe had been repaired); now 30 min, most of which I can be in my office doing something else. So the 3D shim routine (takes about 20 mins) presents a huge time savings, and greatly assists in keeping line shape specs in an on-going manner.
We will also use it on "tough" samples on occasion, when they're in proteo solvents. The most common case would be a protein in H2O:D2O where the water suppression just isn't as good as we're hoping for.
I'll certainly be interested if Bruker has really separated this out and is charging for 3D shimming. It is obviously something we would want, and therefore something I recommend you get, if it's not too expensive. I guarantee that the installation engineer will use it when installing your probe.
Charlie
On 9/25/2018 11:32 AM, Kennedy, Robert wrote:
Hello
College of the Holy Cross is currently writing a grant for a new NMR (400
or 500 MHz). 3D gradient shimming is available from Bruker as an upgrade.
Is 3D gradient shimming the future of shimming?
Does anyone have 3D gradient shimming on a 500 MHz instrument? If so, what
are the added benefits?
Thank you,
Bobby Kennedy
--
Bobby Kennedy
Organic Chemistry Lab Supervisor
Department of Chemistry
College of the Holy Cross
One College Street
Worcester, MA 01610-2395
508-793-2492
Smith Labs 258
--
~~~~~~~~~~~~~~~~~~~~~~~
Charles G. Fry, Ph.D. Tel: (608)262-3182
Director, MR Facility Fax: (608)262-0381
Chem. Dept., 1101 University Ave, Univ. Wisconsin-Madison
Madison, WI 53706 USA email: fry_at_chem.wisc.edu<mailto:fry_at_chem.wisc.edu>
--
==============================================================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive Phone:978 667 9580 ext 5144
Billerica MA 01821 skype_ID: canklin
web: www.bruker.com<http://www.bruker.com> e-mail: clemens.anklin_at_bruker.com<mailto:clemens.anklin_at_bruker.com>
==============================================================
Received on Wed Sep 26 2018 - 01:45:39 MST