RE: AMMRL: 3D Gradient Shimming

From: Shoemaker, Richard <Richard.Shoemaker_at_bruker.com>
Date: Wed, 26 Sep 2018 13:37:49 +0000

Hello Everyone,

Since leaving my position at Colorado I have been doing most of my AMMRL correspondence from my personal email account, but I believe this one is appropriate to respond to from my role at Bruker.

Many thanks to Charlie for responding with the correct information, and I too used Topshim-3D on many occasions with much success on my three Bruker systems in Boulder, CO (in all of the types of situations that Charlie mentioned).

I just want to assure everyone that 3D-Topshim will continue to be an integral part of TopSpin, as it always has been, with no extra charge. Somehow some incorrect information was communicated, and I apologize for that.

Best wishes to everyone from the other side ;o)

-Rich Shoemaker

From: CHARLES G FRY <fry_at_chem.wisc.edu>
Sent: Tuesday, September 25, 2018 12:46 PM
To: Kennedy, Robert <rkennedy_at_holycross.edu>; ammrl_at_ammrl.org
Subject: Re: AMMRL: 3D Gradient Shimming

Hi Bobby,

This is curious. 3D gradient shimming has always in the past been available at no charge in TopSpin. It is definitely a useful feature at any point where the probe has been worked on, and when it is being first installed. It is easy to use, and does a good job. We use it regularly to help us update our saved shims, something we do about monthly. It is from those saved shims that we then do the 1D gradient shimming on the in-between. The key here is that it pulls in good values for the non-axial (non-Z) shims. These can be very hard to manually shim in. Back in the old days, it would take 1-2 days to shim in a new probe. Now it takes maybe an hour. Updating a shim set used to take me an hour or more (especially if the probe had been repaired); now 30 min, most of which I can be in my office doing something else. So the 3D shim routine (takes about 20 mins) presents a huge time savings, and greatly assists in keeping line shape specs in an on-going manner.

We will also use it on "tough" samples on occasion, when they're in proteo solvents. The most common case would be a protein in H2O:D2O where the water suppression just isn't as good as we're hoping for.

I'll certainly be interested if Bruker has really separated this out and is charging for 3D shimming. It is obviously something we would want, and therefore something I recommend you get, if it's not too expensive. I guarantee that the installation engineer will use it when installing your probe.

Charlie



On 9/25/2018 11:32 AM, Kennedy, Robert wrote:

Hello



College of the Holy Cross is currently writing a grant for a new NMR (400

or 500 MHz). 3D gradient shimming is available from Bruker as an upgrade.



Is 3D gradient shimming the future of shimming?



Does anyone have 3D gradient shimming on a 500 MHz instrument? If so, what

are the added benefits?



Thank you,

Bobby Kennedy



--

Bobby Kennedy

Organic Chemistry Lab Supervisor

Department of Chemistry

College of the Holy Cross

One College Street

Worcester, MA 01610-2395

508-793-2492



Smith Labs 258





--

~~~~~~~~~~~~~~~~~~~~~~~

Charles G. Fry, Ph.D. Tel: (608)262-3182

Director, MR Facility Fax: (608)262-0381

Chem. Dept., 1101 University Ave, Univ. Wisconsin-Madison

Madison, WI 53706 USA email: fry_at_chem.wisc.edu<mailto:fry_at_chem.wisc.edu>
Received on Wed Sep 26 2018 - 03:38:05 MST

This archive was generated by hypermail 2.4.0 : Tue Jun 20 2023 - 18:50:45 MST