Dear Victoria,

We had a similar issue in our SampleCase 60, several times it got jammed while
moving between concentric.

Finally the support technician from Bruker discovered that the cause was the cover.
It is made by two metallic layers which slid between them when moving between
concentric position rings. Between these metallic layers there is a polymer that
had degraded and became sticky, causing the system to get stuck from time to time.

We could have had the cover replaced by a new one, but we chose just to remove it
and it has been working wonderfully since then.

Hope that helps. Best wishes,

Gabriel


[Descripción: logoICIQ]
Dr. Gabriel González Ferrer
Responsable RMN / NMR Manager
Institut Catala d'Investigacio Quimica (ICIQ) /
Institute of Chemical Research of Catalonia (ICIQ)
The Barcelona Institute of Science and Technology
Av. Paisos Catalans 16 - 43007 Tarragona (Spain)
Phone +34 977920207 / +34 686669420 - Fax +34 977920224
ggonzalez_at_iciq.es<mailto:ggonzalez_at_iciq.es>


This e-mail contains proprietary information whose content can be protected
by law. If you are not the intended recipient, you may not use, disclose,
distribute, copy or print this e-mail. This confidential, personal and not
transferable e-mail is intended for the addressee only and is legally protected.
Any disclosure, unauthorized use or forwarding, in whole or in part is not
permitted. If you are not the intended recipient of this e-mail message and
its contents, please notify the sender immediately and delete this message
and all its attachments subsequently without reading or record, in whole
or in part. Thank you.

Aquest correu electronic conte una informacio propietaria el contingut del qual
pot estar legalment protegit. Si voste no es el destinatari previst, no pot
usar, divulgar, distribuir, copiar o imprimir aquest correu electronic. Aquest
missatge electronic esta dirigit unicament a la/les direccio/ns indicada/es
anteriorment; el caracter confidencial, personal i intransferible del mateix
esta protegit legalment. Qualsevol revelacio, us o reenviament no autoritzat,
complet o en part, esta prohibit. Si ha rebut aquestmissatge per equivocacio,
notifiqui-ho immediatament a la persona queho ha enviat i esborri el missatge
original juntament amb els seus fitxers annexos sense llegir-lo ni gravar-lo,
totalment o parcialment. Gracies.

Este correo electronico contiene una informacion propietaria cuyocontenido
puede estar legalmente protegido. Si usted no es el destinatario previsto, no puede
usar, divulgar, distribuir, copiar o imprimir este correo electronico. Este
mensaje electronico esta dirigido unicamente a la/s direccion/es indicada/s
anteriormente; el caracter confidencial, personal e intransferible del mismo
esta protegido legalmente. Cualquier revelacion, uso o reenvio no autorizado,
completo o en parte, esta prohibido. Si ha recibido este mensaje por equivocacion,
notifiquelo inmediatamente a la persona que lo ha enviado y borre el mensaje original
junto con sus ficheros anexos sin leerloni grabarlo, total o parcialmente. Gracias.


> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of vbegue via groups.io
> Sent: Friday, June 9, 2023 2:57 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Bruker SampleCase Reliability #Bruker


[Edited Message Follows]
[Reason: edited to correct the number of slots]
All,

I'm wondering if anyone can make any comments about the reliability of the
Bruker SampleCase with 24 slots vs 60 slots. We have both configurations in
our lab and the 60 slot model is having problems with jamming as it moves
between the concentric circles. The 24 slot model only has one ring and does
not encounter this problem. I'm wondering if the jamming issues I'm seeing
in the 60 slot model are just sort of an inherent 'feature' of the increased
alignment requirements generated by the concentric ring design or if I've
just had bad luck with the 60 slot model and other people are finding them
generally reliable.

Thanks,
Victoria Begue



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#191): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/191__;!!PvDODwlR4mBZyAb0!QxuXnvrtLB77KdzLGsgG7Zp0JywuavaX=
WUuCH2VQD-Q-tpLg4qFSAWGIZFRP_SYmAz4c0JIVHOsT1I0NAQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99427463/=
7559972__;!!PvDODwlR4mBZyAb0!QxuXnvrtLB77KdzLGsgG7Zp0JywuavaXWUuCH2VQD-Q-tp=
Lg4qFSAWGIZFRP_SYmAz4c0JIVHOtOs0BEXw$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!QxuXnvrtLB77KdzLGsgG7Zp0JywuavaXWUuCH2=
VQD-Q-tpLg4qFSAWGIZFRP_SYmAz4c0JIVHOs1H5nIXQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






image001.jpg
(image/jpeg attachment: image001.jpg)

Received on Fri Jun 09 2023 - 07:08:11 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST