Hello Victoria,

Just to clarify, the two models of sample-changer are the SampleCase,
which has 24 sample positions, and the SampleCasePlus, which can hold up
to 60-samples. As you said, the 60-sample model has three rings.

We have both models, and we have been quite pleased with them. In general
we have found these to be extremely reliable. They require twice yearly
maintenance which is described in the owners manual. This maintenance
involves cleaning and lubricating the horizontal bar where the sample is
transported, and replacing an o-ring. We do this ourselves, and it takes
about 15-minutes.

We did encounter an issues with one of our SampleCases: we needed to
replace part of the pneumatic control system when an o-ring buried deep
inside the controller failed. I think this failure was the exception,
as neither of the other two systems have experienced this failure.

The SampleCase (24) is the simpler system.  If you expect that room for
24-samples will meet your needs, then I would advise going with that.
That said, we are extremely pleased with the performance of the
SampleCasePlus (60). This is by far the spectrometer with the highest
volume of sample throughput, and it has been working like a charm for 3+ years.

One thing to consider:  we have several users who work with air/water
sensitive samples who use the "J-Young" type NMR tubes. By default,
adding the valve system to an NMR tube makes them too long to work with
either the SampleCase or SampleCasePlus.  We ended up having to replace
the longer J-Young tubes with shorter ones. These have been working fine.

For me, the biggest selling point of the SampleCase and SampleCasePlus
sample changers is that they are located about a meter above the floor.
Users no longer need to climb up and down ladders to insert/remove
samples.  I never realized how important that was until I saw it, and it
is excellent for accessibility.

Cheers,
Bob.

On 2023-06-09 8:31 a.m., vbegue via groups.io wrote:
>
>
> All,
>
> I'm wondering if anyone can make any comments about the reliability of
> the Bruker SampleCase with 45 slots vs 60 slots. We have both
> configurations in our lab and the 60 slot model is having problems
> with jamming as it moves between the concentric circles. The 45 slot
> model only has one ring and does not encounter this problem. I'm u
> wondering if the jamming issues I'm seeing in the 60 slot model are
> just sort of an inherent 'feature' of the increased alignment
> requirements generated by the concentric ring design or if I've just
> had bad luck with the 60 slot model and other people are finding them
> generally reliable.
>
> Thanks,
> Victoria Begue
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------=
-------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon" wampum agreement.




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#190): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/190__;!!PvDODwlR4mBZyAb0!Sq52DgQ5kQ1W1SOq3M1oqkK7POpy2-OW=
BFydHIsSzizx9N58O8TZfsbq6zDtN5iI5U4Qmu4lx0cFpgA1$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99427463/=
7559972__;!!PvDODwlR4mBZyAb0!Sq52DgQ5kQ1W1SOq3M1oqkK7POpy2-OWBFydHIsSzizx9N=
58O8TZfsbq6zDtN5iI5U4Qmu4lx4d7thQl$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!Sq52DgQ5kQ1W1SOq3M1oqkK7POpy2-OWBFydHI=
sSzizx9N58O8TZfsbq6zDtN5iI5U4Qmu4lx9YDcRlW$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 09 2023 - 06:00:05 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST