Re: [AMMRL] Bruker SampleCase 60 – Over-sensitive safety bar / height sensor workarounds?

From: Hasan Celik via groups.io <hcelik=berkeley.edu_at_groups.io>
Date: Mon, 5 Oct 2026 08:44:38 -0700

Hi Ken,

I strongly suggest updating the firmware. There was an update from Bruker
last year that solved this issue for us.
Best,
Hasan

_____________________________
Director, PMRC Core NMR Facility
College of Chemistry
University of California, Berkeley
D35A Hildebrand Hall
Office: (510) 643-9829
Cell: (949) 732-8175

CONFIDENTIALITY NOTICE: This e-mail, including any attachments, is for the
sole use of the intended recipient and may contain information that is
confidential and privileged under state and federal privacy laws. If you
received this e-mail in error, be aware that any unauthorized use,
disclosure, copying or distribution is strictly prohibited. Please contact
the sender immediately and delete/destroy all copies of this message.

On Mon, Oct 5, 2026 at 7:32 AM Kenneth Sharp-Knott via groups.io wrote:

> Hi everyone,
> We are experiencing frequent, spurious autosampler error trips on our
> 60-position SampleCase. Despite user education, users loading or retrieving
> tubes near the transport mechanism frequently brush against the red safety
> / tube-height detection arm.
> Compared to our 24-position units, the mechanism on the 60-position
> carousel feels loose, has weaker return tension, and seems prone to cause
> errors. Even a slight nudge trips the sensor, turns the status light solid
> red, and halts automation until manually reset.
> Has anyone implemented an effective physical or mechanical workaround for
> this (e.g., auxiliary rubber bands, some kind of maintenance to the pivot
> point, or other suggestions) to prevent false trips without bypassing the
> over-length safety check?
> Any insights or field fixes would be greatly appreciated.
> Thanks,
> Ken Sharp-Knott
> Senior Manager, NMR & Analytical Core Facilities
> Department of Chemistry
> Virginia Tech
>
> (540)267-6502 (Cell)
> (540)231-0885 (Office)
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3818): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3818__;!!PvDODwlR4mBZyAb0!SK0IXDrt4uTjS7njEDXpPWVdKcLxdA=
4NAeyW-AfRqOJ64QN3R8OXzmLrQ7kuiNENWFNeCYqJfsoJZbwbQkos0KnUXz_m$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121592293=
/7559972__;!!PvDODwlR4mBZyAb0!SK0IXDrt4uTjS7njEDXpPWVdKcLxdA4NAeyW-AfRqOJ64=
QN3R8OXzmLrQ7kuiNENWFNeCYqJfsoJZbwbQkos0FSLKrwH$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 05 2026 - 08:45:43 MST

This archive was generated by hypermail 2.4.0 : Mon Oct 12 2026 - 13:25:51 MST