Steve,
Thanks for your input. It sure seems like something similar. I found four
Bruker services with 3 of the four running. Bootparam is the name of the
Bruker boot parameter server which probably should be running but crashes
instantly upon being manually started.
At least I have some thing specific to ask Bruker about!
Cheers
Scott
Scott K. Smith, Director, NMR Core Facility
The University of Texas at Austin | Chemistry | 832-294-1928 | utexas.edu
https://www.utexas.edu/
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of steven.clemens via groups.io
> Sent: Monday, September 18, 2023 9:40 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Trouble with TS3.6.5 Win upgrade
I've fought with this a couple times - once when setting up a new Windows 10
workstation. The symptom for me was that the IPSO would stall out at C0 or 00
when booting, meaning that it has not loaded an OS from Diskless.
The cause of the problem was that the service "Bruker Tftp server" was not
running, even though it was set to start automatically. By starting it manually
and then pressing the reset button on the IPSO, I was able to get it working.
I then set the startup behaviour of the service to be "Automatic (Delayed Start)"
and this seems to have solved the problem.
Steve
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#606): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/606__;!!PvDODwlR4mBZyAb0!WACEHipw9-qCbXB_P_txjBvpkcp_TUK8=
n-LFtgxsbztLCwGZWpyF774os2NrTild1MFUmt5ppRHySMbGrFBkcl0X8ma1$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101433329=
/7559972__;!!PvDODwlR4mBZyAb0!WACEHipw9-qCbXB_P_txjBvpkcp_TUK8n-LFtgxsbztLC=
wGZWpyF774os2NrTild1MFUmt5ppRHySMbGrFBkcnIciVkS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Sep 18 2023 - 14:46:38 MST