Hi Ravi,
The best one is K2PtCl6 / Na2PtCl6 (0 ppm) or K2PtCl4(~-1600 ppm).
Cisplatin has a low water-solubility. The soluble part of cisplatin can
dissociate and release chlorides in water.
Yun
On Wed, Sep 6, 2023 at 1:27 PM Garimella, Ravi <rgarimel_at_odu.edu> w=
rote:
> Hi ,
>
> Is Cisplatin a good reference compound for setting up Pt195?
>
> Thanks,
>
> -Best,
>
> -Ravi Garimella
>
> (Ravindranath Garimella PhD)
> NMR Manager
> College of Sciences Major Instrumentation Cluster (COSMIC) Facility
> 1104 Oceanography and Physical SCIENCES Bidg.,
> 4600 Elkhorn Ave.
> Norfolk, VA 23529.
> Email : rgarimel_at_odu.edu
> Ph : (757)683-3434
>
>
>
>
--
Yun Qu, Ph.D
Director of NMR Center
Department of Chemistry
Virginia Commonwealth University
1001 W. Main St.
Richmond, VA 23284
Phone: 804-655-0998
FAX: 804-828-8599
https://chemistry.vcu.edu/research/facilities/the-nuclear-magnetic-resonance-center/
-=-=-=-=-=-=-=-=-=-=-=-