Re: [AMMRL] upgrade to TS 2.1 pl6 --> TS 3.6.5 raw check no signal on gradient shimming

From: Michael Groves <m.groves_at_utah.edu>
Date: Sat, 2 Sep 2023 01:06:34 +0000

I agree—it does sound like the routing is messed up. Can you send a
screenshot of your edasp window with the “show probe wiring”
box checked? 2H should go through the 5W amplifier—the 100W amp is
connected to a BB preamp that has a 2H stop filter.

Cheers,
Mike

On Sep 1, 2023, at 6:58 PM, Darcy Burns <darcy.burns_at_utoronto.ca> wrote:


Well folks,

I’ve seemingly bricked out 2h gradient performance when attempting
to upgrade TS 2.1 to TS 3.6.5 (CentOS / AV III nano). In this case we are
receiving a topshim raw check - no signal at all

In a nutshell:
1. we are able to acquire a 1D 2H (zg2h)
2. we are able to topshim 1h
3. We are able to make a new shimmap
4. 1D 1H and 1D 13C both are OK
5. I can manually shim to lineshape spec

In this case, we installed and ran nmrrestore to bringing out tar file. Then
cf / ii / ii restart to clear any issues. Then checked prosol table and=
edited with previous values accordingly.

At first we were not able to get a good shim response (shimming iterated many
times and set to poor values, complained of poor signal:noise, which in
retrospect may have been related a conversion from dB to dBm in prosol).
Still, we did not get a raw check no signal at all error.


We then though (perhaps unwisely) that there was an issue with the shimmap
and performed a ‘topshim initial’.
à at this point we were no longer able to get a 2H response in topshim.
1D 2H is OK, so is 1h gradient shimming.
à seems like the routing is messed up


Routing is as follows:
1H à SGU2 à 60W amp
2H à SGU1 à 100W amp
5W amp is available, but unused.



I have no idea what to do; we reverted to the older computer running TS 2.1,
ran an nmrrestore and went through requisite cf’s; all to no avail.

Any advice out there?

cheerios,
Darcy

_________________________________
Darcy C. Burns, Ph.D.
Manager, CSICOMP NMR Facility

NEW BOOK! Optimizing NMR Methods for Structure Elucidation (Burns and Reyn=
olds)

University of Toronto
80 St. George Street
Toronto, ON
M5S 3H6
ph: (416)-978-0448
fax: (416)-978-8775
https://sites.chem.utoronto.ca/csicomp/__


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#528): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/528__;!!PvDODwlR4mBZyAb0!Q1A1g-_Ny0FnskgrDuJ1bR3PYjQFh4mz=
W0ZsGqgdnBnzT88KXlgV4ew5fGXdX70pEEjRnu7wmx8cJFXbtQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101107231=
/7559972__;!!PvDODwlR4mBZyAb0!Q1A1g-_Ny0FnskgrDuJ1bR3PYjQFh4mzW0ZsGqgdnBnzT=
88KXlgV4ew5fGXdX70pEEjRnu7wmx-OU1EuMA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




From owner-ammrl_at_ammrl.groups.io Fr, 1 Sep 2023 18:13:21
Received: by 2002:a05:7022:2649:b0:6a:a843:e96f with SMTP id er9csp681827dlb;
        Fri, 1 Sep 2023 18:13:21 -0700 (PDT)
X-Google-Smtp-Source: AGHT+IGg/NEyNNMmYlVdv93g1zPf59rWGAdKOCwX7FzjLXOQOcSz9vfLYvcC+wR4QIeb9doo2jEP
X-Received: by 2002:a05:6a00:24c9:b0:68c:b30:8fcc with SMTP id d9-20020a056a0024c900b0068c0b308fccmr4611046pfv.20.1693617201307;
        Fri, 01 Sep 2023 18:13:21 -0700 (PDT)
Return-Path: <bounce+115296+529+7559972+12177410_at_groups.io>
Received: from mx0a-007a6701.pphosted.com (mx0a-007a6701.pphosted.com. [205.220.169.170])
        by mx.google.com with ESMTPS id cn10-20020a056a00340a00b00688452c12e1si3771675pfb.0.2023.09.01.18.13.20
        for <walt_at_hawaii.edu>
        (version=TLS1_2 cipher=ECDHE-RSA-AES128-GCM-SHA256 bits=128/128);
        Fri, 01 Sep 2023 18:13:21 -0700 (PDT)
Received-SPF: pass (google.com: domain of bounce+115296+529+7559972+12177410_at_groups.io designates 66.175.222.108 as permitted sender) client-ip=66.175.222.108;
Authentication-Results: mx.google.com;
       dkim=neutral (body hash did not verify) header.i=_at_groups.io header.s=20140610 header.b="nxkdHZ/G";
       arc=fail (body hash mismatch);
       spf=pass (google.com: domain of bounce+115296+529+7559972+12177410_at_groups.io designates 66.175.222.108 as permitted sender) smtp.mailfrom=bounce+115296+529+7559972+12177410_at_groups.io
Received: from pps.filterd (m0313024.ppops.net [127.0.0.1])
        by mx0a-007a6701.pphosted.com (8.17.1.19/8.17.1.19) with ESMTP id 381M8HoQ026018
        for <walt_at_hawaii.edu>; Fri, 1 Sep 2023 15:13:18 -1000
Authentication-Results: ppops.net;
        spf=pass smtp.mailfrom=bounce+115296+529+7559972+12177410_at_groups.io;
        dkim=pass header.d=groups.io header.s=20140610;
        dmarc=fail header.from=pitt.edu
Received: from mail02.groups.io (mail02.groups.io [66.175.222.108])
        by mx0a-007a6701.pphosted.com (PPS) with ESMTPS id 3su6j7ccxv-1
        (version=TLSv1.2 cipher=ECDHE-RSA-AES128-GCM-SHA256 bits=128 verify=NOT)
        for <walt_at_hawaii.edu>; Fri, 01 Sep 2023 15:13:17 -1000
DKIM-Signature: a=rsa-sha256; bh=otQOv7k6/2jiS84sRJMZN83BMN3fN2rywULlDmEQySU=;
 c=relaxed/simple; d=groups.io;
 h=ARC-Seal:ARC-Message-Signature:ARC-Authentication-Results:From:To:Subject:Thread-Topic:Thread-Index:Date:Message-ID:References:In-Reply-To:Accept-Language:MIME-Version:Precedence:List-Subscribe:List-Help:Sender:List-Id:Mailing-List:Delivered-To:Reply-To:List-Unsubscribe-Post:List-Unsubscribe:Content-Language:Content-Type;
 s=20140610; t=1693617197; v=1;
 b=nxkdHZ/GvzGQ66rbACsmz2PWsclaCTTIDtnRpoTpXI7vizPXaUKGpdRCnnt9ELaRZr3fXmyJ
 qIKmy1GOo9hrKDGcuYICdzvD1Fq4wNOeBX+Px+QT9pC8oGgU26LfUY6/KjblA8ju1jnNDwzYoGL
 Iu51pE28Wd/V08qkg8M9BqsM=
X-Received: by 127.0.0.2 with SMTP id UdWQYY7561307xX6uXvP2M7M; Fri, 01 Sep 2023 18:13:17 -0700
X-Received: from NAM10-DM6-obe.outbound.protection.outlook.com (NAM10-DM6-obe.outbound.protection.outlook.com [40.107.93.40])
 by mx.groups.io with SMTP id smtpd.web11.1101.1693617194310908598
 for <main_at_ammrl.groups.io>;
 Fri, 01 Sep 2023 18:13:16 -0700
ARC-Seal: i=1; a=rsa-sha256; s=arcselector9901; d=microsoft.com; cv=none;
 b=f+5uczvVzQen8Vur+KJPd+ghAMEJlvi6qbByXULt/qJ55ijnOXoD94WFm/saGfozA4sOnzmh00vWj6NhlyXBG5fZLvYtySFswHp3KvoZBZeZYbfNYNmyf+bWbw2vQkDjMK/2Bi6r6GjfoUzIBcS2tFmqaP67Y2JMlOnMRrHSJNhhJjtd7eZoe+4rE+8dOW9d46/gs/8Vi9yRh0aDOARjEmQWXyfbjcl8KOh6RA48vyZptaVW6QH5URITb9sV7IfPfXaegrowEVNQ0jDhFjjwklsresPMG3QUBY2kRS02XnCpfB+/4f695L7jnqqkEzPxnR5gGmsqMI7vLyUC02sTZA==
ARC-Message-Signature: i=1; a=rsa-sha256; c=relaxed/relaxed; d=microsoft.com;
 s=arcselector9901;
 h=From:Date:Subject:Message-ID:Content-Type:MIME-Version:X-MS-Exchange-AntiSpam-MessageData-ChunkCount:X-MS-Exchange-AntiSpam-MessageData-0:X-MS-Exchange-AntiSpam-MessageData-1;
 bh=NbqRiCWgbxKl5CyQBo+SZXAHmqjirs1GD4uMgvxTgrY=;
 b=dka9XYwSw7d7DIptpqqzxUameDAB1TZ+LZSIFZ56MjB/YubKDA3ZfWB7F7eZJdDIZFqEHiAcc2dJpcDYbwecALyRkyv+9QCyg6Qk0abaXTkzr+igDoaldqvKmRkepxiZspOzEB73JX5kYXIgzdt5otwzVD/Up+oM644682NtUmmNG7TV235kr90N1QF9inzXkj580kYu6hRo0VuOLKrwq7htZXv0Y6xfaUZHnOqYybQ3bnBTLPNNVNGG9GCcQHTssWE//NVwOPCg32JWUjcJtffTAaplcqOcX8CBGTDXN9uVctY/nLR5Og0yudFr4TxjVuLvHJiNdjwQAT+Mlnwi8g==
ARC-Authentication-Results: i=1; mx.microsoft.com 1; spf=pass
 smtp.mailfrom=pitt.edu; dmarc=pass action=none header.from=pitt.edu;
 dkim=pass header.d=pitt.edu; arc=none
X-Received: from SN4PR04MB8320.namprd04.prod.outlook.com (2603:10b6:806:1eb::7)
 by DM6PR04MB6762.namprd04.prod.outlook.com (2603:10b6:5:24f::23) with
 Microsoft SMTP Server (version=TLS1_2,
 cipher=TLS_ECDHE_RSA_WITH_AES_256_GCM_SHA384) id 15.20.6768.9; Sat, 2 Sep
 2023 01:13:10 +0000
X-Received: from SN4PR04MB8320.namprd04.prod.outlook.com
 ([fe80::71ee:380e:9392:6332]) by SN4PR04MB8320.namprd04.prod.outlook.com
 ([fe80::71ee:380e:9392:6332%7]) with mapi id 15.20.6768.017; Sat, 2 Sep 2023
 01:13:09 +0000
From: "Delk, Michael Joseph" <mid15_at_pitt.edu>
To: "main_at_ammrl.groups.io" <main_at_ammrl.groups.io>
Subject: Re: [AMMRL] upgrade to TS 2.1 pl6 --> TS 3.6.5 raw check no signal on gradient shimming
Thread-Topic: [AMMRL] upgrade to TS 2.1 pl6 --> TS 3.6.5 raw check no signal
 on gradient shimming
Thread-Index: AdndFKbh9MrjTG8UQmaR7ABC1gnN5wAJRBw4AAArZzU=
Date: Sat, 2 Sep 2023 01:13:09 +0000
Message-ID:
 <SN4PR04MB83206DF0DEBCA75D5E84FA61DDEBA_at_SN4PR04MB8320.namprd04.prod.outlook.com>
References:
 <YT1PR01MB36763C66B7440104AC69849EEBE4A_at_YT1PR01MB3676.CANPRD01.PROD.OUTLOOK.COM>
 <A9E7BF96-BB43-48B5-9B57-27E795B2444E_at_utah.edu>
In-Reply-To: <A9E7BF96-BB43-48B5-9B57-27E795B2444E_at_utah.edu>
Accept-Language: en-US
X-MS-Has-Attach:
X-MS-TNEF-Correlator:
x-ms-publictraffictype: Email
x-ms-traffictypediagnostic: SN4PR04MB8320:EE_|DM6PR04MB6762:EE_
x-ms-office365-filtering-correlation-id: 3c6b2bda-312d-47a8-ea29-08dbab51c588
x-ms-exchange-senderadcheck: 1
x-ms-exchange-antispam-relay: 0
x-microsoft-antispam-message-info:
 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
x-ms-exchange-antispam-messagedata-chunkcount: 1
x-ms-exchange-antispam-messagedata-0:
 =?utf-8?B?c0d4ZlFXOVZkVStwbWJaS0M3WU9BVUVwZDBGWFl6VFRJdVFJeTQ0ZDVCUDRi?=
 =?utf-8?B?bXo2WEFTMitNc3ZmSkJIWndWTVROb1NJOWdkVVE1eC8vQ09xRXBURVVWdVNV?=
 =?utf-8?B?MlZSQXhLT1VYR0hVZlYrT25UTEQ4dDNkY0tic0s2V2Z0S3ZoWVJOb1lDYktU?=
 =?utf-8?B?TXB5MkFaTzFYaVpOVk5NdXAzUXdQQ3NPTnNuUDRSOTRGd1pENHhuM2dpcDJS?=
 =?utf-8?B?WHYxZHZQaEF1OWtPdHVLdndaZzIvUUdUSjN5T0lTQmFSSHVob1hBQkpMemlR?=
 =?utf-8?B?VUo3bkp2Y1BKdEFLbXNiVzUvdXd1cXRaNEhDZ2VwOU9jWVlZN2hJY1F3Qi9l?=
 =?utf-8?B?UDlHd3Y0eVJTT3V6Sk5jeWduT1RzUXJCalVQKzZ4WHVGK2ZmUXQ5VWxXem1V?=
 =?utf-8?B?V1hzeUs2clFZU2R3ckV0ejE3OXRwNDduaHFPZUJ2aTRBN2FxSnpFOVY4UjJh?=
 =?utf-8?B?M1hwQTdxNkhUdkhPekcvUTNNb3pxVnFmYThNL2RiODE4dW80S2JUcXgwSU91?=
 =?utf-8?B?SEZNUGNHcHNSRkc2eC9KeEt1bWdQMG9VVFlTMU9VdHF6TnZUTGNNb1QvV0VR?=
 =?utf-8?B?dTVHcyt6MEFSOVVPTnIwdXMwT3NJVldVeDFPZzVZNnpSMEhRemxPVnZhTWZs?=
 =?utf-8?B?bnlCVDJOTms5Y3k3OWZIdUN1TzU2NW5VaDFFNVc5V0VrT2NZcGFuaUtUZkNM?=
 =?utf-8?B?Y3R5ajNjSHRabGE5WFMzVXU0TnBJSjhvd203Z1lvL3RKeEJ0T0ZCeHkzcGFL?=
 =?utf-8?B?d2t0NUV5bjdQa2dOYTFWSW12T2E0VXZ0b1JxZ2hwSGhxZm9xamlub3NmazRj?=
 =?utf-8?B?RGZscWpRaW9yaUZIV1FyeHQwOFR3bHNod3Qvc1BDQ2lvYVJLb0I4K0RsWEhh?=
 =?utf-8?B?Y1RzajBiQ1JMaVd4TEw2SGhYY3pCUEREWnJhbXFlWk9WT1ZIQWZFM3ZhMGwy?=
 =?utf-8?B?S0VpZnhIVjByYXBKdjZKdDE0U2VodDdOMGVLK1hhYWFsVTIzR05qRm9WNGlj?=
 =?utf-8?B?cUJyQkQ2UVdNUkVMUzVVY1JGbUZIQXV3citTMDQ1Y01RejhTRE9iN3NWSnNG?=
 =?utf-8?B?SWZ6RXNWY2UvSkNSZzlOZ0h1Uk5NRmFEUGxxNDFZZ0MxMlVaa244VTk2cjA1?=
 =?utf-8?B?cmM2ZkQvNjFLTUVlcDRoR0pzSFNpcWgzb0VYSlJlWVF6TThSWFhuMm1vRkUw?=
 =?utf-8?B?azZGWUJwaDNTRCtvNlNWTEhmbWE5MUt3MEE0TUFkL1l3VHNvZkNtd1dacDRE?=
 =?utf-8?B?SU1GSnZGWUR3ZFFxVVlNbjNEdVc3cVQwL2tpcHgxQnZsRmZ6dTJEWG5ON0Rx?=
 =?utf-8?B?NW1lakd6cjNmSDlXRStpMTk4cDdPRDlIUlZUWmtuN05kWlRKdm9PRG9TWjl6?=
 =?utf-8?B?YmNlWXByc2VlSkJ2MTNYa3JKdk9NT1VLMDFmVVlpTFhPd3NFZUNFSnZVYTYx?=
 =?utf-8?B?eXd3Qzd4akxnaTFwZSt0OENLYlF4OVYzTkR0alNZOGQ0RnVUWFJMRk9TNW8v?=
 =?utf-8?B?a1JWVUZZL0NHamVPK0pUNmU5S1E2UTFLS0JSaThuQUFmQWxKY1o2aVRrbmVs?=
 =?utf-8?B?RzZEYXhZeWwwandBNUc4dk4ySFpsSkFLZ1E1RmgzdjB3RmZjbDdvR2lVTUxZ?=
 =?utf-8?B?M29xMXU2L3BQQjRjSjlXRnJJZXVyTG81ckRZTjRpUG9ia0NIeVNIZy9nSnkr?=
 =?utf-8?B?RWx2Y2l5NVR4aEVYUTc3aUxHeEVmZGMwVlpwM2RlV002U1FRQnZGM1d5dTRC?=
 =?utf-8?B?SXgzeEdmWk9UY1ZIZU9tdHFaUHNya29IM0d3TTNuRWgzclNuamZaSXJWNlJz?=
 =?utf-8?B?ZHRFQldLQWNxTUtIOS9Od0IrK0h3UzNSNUxtbmZRcVg3NUJzbnJMSGsyREE4?=
 =?utf-8?B?Vld4TVozOFI0SDdSWDhDL3hsSHExbE5rRVhnVGJvaUZYdVlPK0duOEwyZ1gx?=
 =?utf-8?B?Z1FZajdBcm41VUVDYjdnYWt2QlAvZSs4alVQQmZOcHVDZXg4cmhVYTdOcDd1?=
 =?utf-8?B?SmpubEdUencrc1RmRFVpR1NMWWw2eVhIYldPT3I0OVpsTG9QTEVRZlpzNVRy?=
 =?utf-8?B?Q0txWW80QzdFcWZBQVBIeEtLcmQydEZleWJDeXk3ZW80dFRJdU15bnNhVFcy?=
 =?utf-8?Q?HXR244JPYxQmOyT0zFo6PoMMc?=
X-OriginatorOrg: pitt.edu
X-MS-Exchange-CrossTenant-AuthAs: Internal
X-MS-Exchange-CrossTenant-AuthSource: SN4PR04MB8320.namprd04.prod.outlook.com
X-MS-Exchange-CrossTenant-Network-Message-Id: 3c6b2bda-312d-47a8-ea29-08dbab51c588
X-MS-Exchange-CrossTenant-originalarrivaltime: 02 Sep 2023 01:13:09.7461
 (UTC)
X-MS-Exchange-CrossTenant-fromentityheader: Hosted
X-MS-Exchange-CrossTenant-id: 9ef9f489-e0a0-4eeb-87cc-3a526112fd0d
X-MS-Exchange-CrossTenant-mailboxtype: HOSTED
X-MS-Exchange-CrossTenant-userprincipalname: 2ZB7D4p38+mj7BIUZtf5C8/7VRBjrT0LYLqd/AdcRE23T090KHOdnaje8vFOJJVK
X-MS-Exchange-Transport-CrossTenantHeadersStamped: DM6PR04MB6762
Precedence: Bulk
List-Subscribe: <mailto:main+subscribe_at_ammrl.groups.io>
List-Help: <mailto:main+help_at_ammrl.groups.io>
Sender: main_at_ammrl.groups.io
List-Id: <main.ammrl.groups.io>
Mailing-List: list main_at_ammrl.groups.io; contact main+owner_at_ammrl.groups.io
Delivered-To: mailing list main_at_ammrl.groups.io
Reply-To: main_at_ammrl.groups.io
List-Unsubscribe-Post: List-Unsubscribe=One-Click
List-Unsubscribe: <https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/plugh>
X-Gm-Message-State: b4CgRmuezZeFUF0NHaGKXoSnx7559972AA=
Content-Language: en-US
Content-Type: multipart/alternative;
        boundary="_000_SN4PR04MB83206DF0DEBCA75D5E84FA61DDEBASN4PR04MB8320namp_"
X-Proofpoint-GUID: G6fXZFkG-YO83g2NnH5i3ux02meeIRo3
X-Proofpoint-ORIG-GUID: G6fXZFkG-YO83g2NnH5i3ux02meeIRo3
X-CLX-Shades: MLX
X-CLX-Response: 1TFkXGB8bEQpMehccGhEKWUQXYEVuQ0tHQ11rHEwRClhYF25uXVMaZ0JuZHt uEQp4ThdobGdBEktpchh9cxEKeEsXbm5dUxpnQm5ke24RCnlMF2VbEhx6e2EaTRtyEQpDSBcHGx 8eEQpDWRcHGBIcEQpDSRcaBBoaGhEKWU0XZ2ZyEQpZSRcacRoQGncGGBkbcRgaGRAadwYYGgYaE
 QpZXhdobmYRCklGF19EQ1xPWFlDXlNCS11LQ0N1QkVZXk9OEQpJRxd4T00RCkNOF20cTHJwbEFt B3NlEhlNGGREYh9DGV9SGhhHT09jeEUZEQpYXBcfBBoEGRgdBRsaBBsaGgQbGR4EGR8QGx4aHxo RCl5ZF0lhbm5jEQpNXBceGhoRCkxaF2xtUk1dEQpNThdoEQpMRhdea2sRCkJPF20eUkcBGHBIZn
 JrEQpDWhccHAQbHR8EGBgYBBsaEhEKQl4XGxEKRF4XGBEKQkYXYEVuQ0tHQ11rHEwRCkJHF2JdX hpwR3NOZXpBEQpCRRduZGlBGRN9f11aHBEKQk4XYgFjHm5tU3obSUURCkJMF2R9BW1yfntCT2ZB EQpCbBdga1hdf2Zbf0NQUhEKQkAXYGtMXUZ4el9jRwERCkJYF25waGNYe258G1NMEQpaWBccEQp
 5QxdrXEAYTG5pQmlhaBEKWUsXGxgeExsRClpLFx4cEhsRCnBnF2YBRx5PHWJTYx9JEB4SEQpwaB djGHpbRHsBbQVuRBAaEQpwaBdpYxIfWm1sXW1mQRAaEQpwaBdrW2hzAV4YUk0cTBAaEQpwaBdtB RJQHnxna15/RhAaEQpwaBdhZUVDZ2VIHWliYxAZGhEKcGgXZG5JQlkfQhhAH28QGhEKcGgXZ1hy
 aHJAHn5OTHkQGhEKcGgXYGNFWnJveU96Xm0QGhEKcGcXYlBzX0JrQnB8fQEQGhEKcGcXenN5T39 IGnBnQ1AQGhEKcH0XYURNUm9hARNLXWAQGRoRCnB9F2hPfF5vb1hiHUdrEBkaEQpwfRdhQkhQYR xhaGR9SxAaEQpwfRdiXGBkYGlMQRlnYhAaEQpwZxdlZllsXkdoH09ffRAeEhEKcH0XZ2wfUH9DT
 kZTQwEQGhEKcGcXbGxdQU4dTh1eT2UQGhEKcH0XbHlyY30SGF5nbWEQGRoRCnB9F2x4fB5bUhpl XhxiEBkaEQpwfRdhbkgBWnN8bhlhWxAaEQpwfxd6f0dTYxhcGkxIRBAbGRMRCnBfF2NffGsFc01 kaBIBEBkaEQpwfxd6HH5NckZ9S1B9EhAHGx8RCnBfF2JcHFxJTAVtR19IEBkaEQpwfxdsGVkBbW
 QSZkUZfhAbHR8RCnBfF2ZaW1xvR0tuck5TEB4SEQpwXxdsXHsSbW8TT3AfRRAbEhgRCnBfF2h8T 2BQbWhrWR5jEB4SEQpwXxdseHweW1IaZV4cYhAZGhEKcF8XawUaaH1dUFJgcBMQHhIRCnBsF2dj ehwbcmZdaGhQEBoRCm1+FxoRClhNF0sRIA==
MIME-Version: 1.0
X-Proofpoint-Virus-Version: vendor=baseguard
 engine=ICAP:2.0.267,Aquarius:18.0.957,Hydra:6.0.601,FMLib:17.11.176.26
 definitions=2023-09-02_01,2023-08-31_01,2023-05-22_02
X-Proofpoint-Spam-Details: rule=inbound_notspam policy=inbound score=0 spamscore=0 malwarescore=0
 adultscore=0 bulkscore=0 mlxlogscore=999 priorityscore=60 phishscore=0
 mlxscore=0 lowpriorityscore=0 impostorscore=0 clxscore=251 suspectscore=0
 classifier=spam adjust=0 reason=mlx scancount=1 engine=8.12.0-2308100000
 definitions=main-2309020010 domainage_hfrom=12491 domainage_replyto=4681

--_000_SN4PR04MB83206DF0DEBCA75D5E84FA61DDEBASN4PR04MB8320namp_
Content-Type: text/plain; charset="utf-8"
Content-Transfer-Encoding: quoted-printable

If you didn=E2=80=99t do an expinstall after all the changes can you call in a
standard Bruker parameter set and start again from there? Then read in whatever
gradshim1d2h standard set?

Mike Delk
University of Pittsburgh
School of Medicine
Department of Structural Biology
412.370.0066
________________________________
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Michael Groves via groups.io
> Sent: Friday, September 1, 2023 9:06:34 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] upgrade to TS 2.1 pl6 --> TS 3.6.5 raw check no signal on gradient shimming

I agree=E2=80=94it does sound like the routing is messed up. Can you send a
screenshot of your edasp window with the =E2=80=9Cshow probe wiring=E2=80=9D
box checked? 2H should go through the 5W amplifier=E2=80=94the 100W amp is
connected to a BB preamp that has a 2H stop filter.

Cheers,
Mike


On Sep 1, 2023, at 6:58 PM, Darcy Burns <darcy.burns_at_utoronto.ca> wrote:



Well folks,

I=E2=80=99ve seemingly bricked out 2h gradient performance when attempting
to upgrade TS 2.1 to TS 3.6.5 (CentOS / AV III nano). In this case we are
receiving a topshim raw check - no signal at all

In a nutshell:
1. we are able to acquire a 1D 2H (zg2h)

2. we are able to topshim 1h
3. We are able to make a new shimmap
4. 1D 1H and 1D 13C both are OK
5. I can manually shim to lineshape spec

In this case, we installed and ran nmrrestore to bringing out tar file. Then
cf / ii / ii restart to clear any issues. Then checked prosol table and
edited with previous values accordingly.

At first we were not able to get a good shim response (shimming iterated many
times and set to poor values, complained of poor signal:noise, which in
retrospect may have been related a conversion from dB to dBm in prosol).
Still, we did not get a raw check no signal at all error.


We then though (perhaps unwisely) that there was an issue with the shimmap
and performed a =E2=80=98topshim initial=E2=80=99.
=C3=A0 at this point we were no longer able to get a 2H response in topshim.
1D 2H is OK, so is 1h gradient shimming.
=C3=A0 seems like the routing is messed up


Routing is as follows:
1H =C3=A0 SGU2 =C3=A0 60W amp
2H =C3=A0 SGU1 =C3=A0 100W amp
5W amp is available, but unused.

I have no idea what to do; we reverted to the older computer running TS 2.1,
ran an nmrrestore and went through requisite cf=E2=80=99s; all to no avail.

Any advice out there?

cheerios,

Darcy



_________________________________
Darcy C. Burns, Ph.D.
Manager, CSICOMP NMR Facility



NEW BOOK! Optimizing NMR Methods for Structure Elucidation (Burns and Reyn=
olds)

University of Toronto
80 St. George Street
Toronto, ON
M5S 3H6
ph: (416)-978-0448
fax: (416)-978-8775

https://sites.chem.utoronto.ca/csicomp/


-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#529): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/529__;!!PvDODwlR4mBZyAb0!X4pZg1VgjczVo7mXiIxE_YflLfRhMwML=
n_m8AKHwfgNXyEIl0QeSKpOqjiwxi49ZOE0O6mn_Gc3M7Q$=20
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101107231=
/7559972__;!!PvDODwlR4mBZyAb0!X4pZg1VgjczVo7mXiIxE_YflLfRhMwMLn_m8AKHwfgNXy=
EIl0QeSKpOqjiwxi49ZOE0O6mnYW_LQtA$=20
Group Owner: main+owner_at_ammrl.groups.io
-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-



--_000_SN4PR04MB83206DF0DEBCA75D5E84FA61DDEBASN4PR04MB8320namp_
Content-Type: text/html; charset="utf-8"
Content-Transfer-Encoding: quoted-printable

<html><head>
<meta http-equiv=3D"Content-Type" content=3D"text/html; charset=3Dutf-8"/>
</head>
<body>
<div>
<div>
<div dir=3D"ltr">If you didn=E2=80=99t do an expinstall after all the chang=
es can you call in a standard Bruker parameter set and start again from the=
re? Then read in whatever gradshim1d2h standard set?</div>
</div>
<div id=3D"ms-outlook-mobile-signature">
<div><br/>
</div>
<div dir=3D"ltr">
<div dir=3D"ltr">Mike Delk</div>
<div dir=3D"ltr">University of Pittsburgh=C2=A0</div>
<div dir=3D"ltr">School of Medicine</div>
<div dir=3D"ltr">Department of Structural Biology</div>
<div dir=3D"ltr">412.370.0066</div>
</div>
</div>
</div>
<hr style=3D"display:inline-block; width:98%;" tabindex=3D"-1"/>
<div id=3D"divRplyFwdMsg" dir=3D"ltr"><defangedfont face=3D"Calibri, sans-s=
erif" style=3D"" color=3D"#000000"><b>From:</b> main_at_ammrl.groups.io &lt;ma=
in_at_ammrl.groups.io&gt; on behalf of Michael Groves via groups.io &lt;m.grov=
es=3Dutah.edu_at_groups.io&gt;<br/>
<b>Sent:</b> Friday, September 1, 2023 9:06:34 PM<br/>
<b>To:</b> main_at_ammrl.groups.io &lt;main_at_ammrl.groups.io&gt;<br/>
<b>Subject:</b> Re: [AMMRL] upgrade to TS 2.1 pl6 --&gt; TS 3.6.5 raw check=
 no signal on gradient shimming</defangedfont>
<div>=C2=A0</div>
</div>
<div dir=3D"auto">I agree=E2=80=94it does sound like the routing is messed =
up. Can you send a screenshot of your edasp window with the =E2=80=9Cshow p=
robe wiring=E2=80=9D box checked? =C2=A02H should go through the 5W amplifi=
er=E2=80=94the 100W amp is connected to a BB preamp that has a 2H stop
 filter.=C2=A0
<div><br/>
</div>
<div>Cheers,</div>
<div>Mike<br/>
<br/>
<div dir=3D"ltr"></div>
<div dir=3D"ltr"><br/>
<blockquote type=3D"cite">On Sep 1, 2023, at 6:58 PM, Darcy Burns &lt;darcy=
.burns_at_utoronto.ca&gt; wrote:<br/>
<br/>
</blockquote>
</div>
<blockquote type=3D"cite">
<div dir=3D"ltr">=EF=BB=BF
<meta name=3D"Generator" content=3D"Microsoft Word 15 (filtered medium)"/>
 <style>
<!--
_at_font-face
        {font-family:Wingdings}
_at_font-face
        {font-family:"Cambria Math"}
_at_font-face
        {font-family:Calibri}
_at_font-face
        {font-family:Tahoma}
p.x_MsoNormal, li.x_MsoNormal, div.x_MsoNormal
        {margin:0cm;
        font-size:11pt;
        font-family:Calibri,sans-serif}
span.x_EmailStyle17
        {font-family:Calibri,sans-serif;
        color:windowtext}
.x_MsoChpDefault
        {font-family:Calibri,sans-serif}
_at_page WordSection1
        {margin:72pt}
div.x_WordSection1
        {}
-->
</style>
<div class=3D"x_WordSection1">
<p class=3D"x_MsoNormal">Well folks, <br/>
<br/>
I=E2=80=99ve seemingly bricked out 2h gradient performance when attempting =
to upgrade TS 2.1 to TS 3.6.5 (CentOS / AV III nano).=C2=A0 In this case we=
 are receiving a topshim raw check =C2=A0- no signal at all<br/>
<br/>
In a nutshell:<br/>
1. we are able to acquire a 1D 2H (zg2h)</p>
<p class=3D"x_MsoNormal">2. we are able to topshim 1h<br/>
3. We are able to make a new shimmap<br/>
4. 1D 1H and 1D 13C both are OK<br/>
5. I can manually shim to lineshape spec<br/>
<br/>
In this case, we installed and ran nmrrestore to bringing out tar file.=C2=
=A0 Then cf / ii / ii restart to clear any issues.=C2=A0 Then checked proso=
l table and edited with previous values accordingly.<br/>
<br/>
At first we were not able to get a good shim response (shimming iterated ma=
ny times and set to poor values, complained of poor signal:noise, which in =
retrospect may have been related a conversion from dB to dBm in prosol).=C2=
=A0 Still, we did not get a raw check
 no signal at all error.<br/>
<br/>
<br/>
We then though (perhaps unwisely) that there was an issue with the shimmap =
and performed a =E2=80=98topshim initial=E2=80=99.<br/>
<span style=3D"">=C3=A0</span> at this point we were no longer able to get =
a 2H response in topshim.=C2=A0 1D 2H is OK, so is 1h gradient shimming.<br=
/>
<span style=3D"">=C3=A0</span> seems like the routing is messed up<br/>
<br/>
<br/>
Routing is as follows:<br/>
1H <span style=3D"">=C3=A0</span> SGU2 <span style=3D"">=C3=A0</span> 60W a=
mp<br/>
2H <span style=3D"">=C3=A0</span> SGU1 <span style=3D"">=C3=A0</span> 100W =
amp<br/>
5W amp is available, but unused.<br/>
<br/>
<br/>
<br/>
I have no idea what to do;=C2=A0 we reverted to the older computer running =
TS 2.1, ran an nmrrestore and went through requisite cf=E2=80=99s;=C2=A0 al=
l to no avail.=C2=A0
<br/>
<br/>
Any advice out there?<br/>
<br/>
cheerios, </p>
<p class=3D"x_MsoNormal">Darcy</p>
<p class=3D"x_MsoNormal">=C2=A0</p>
<div id=3D"x_divtagdefaultwrapper">
<div>
<div>
<div>
<p class=3D"x_MsoNormal"><span style=3D"">_________________________________=
<br/>
Darcy C. Burns, Ph.D.<br/>
Manager, CSICOMP NMR Facility</span></p>
</div>
<div>
<p class=3D"x_MsoNormal"><span style=3D"">=C2=A0</span></p>
</div>
<div>
<p class=3D"x_MsoNormal"><span style=3D"">NEW BOOK!=C2=A0 Optimizing NMR Me=
thods for Structure Elucidation (Burns and Reynolds)</span><span style=3D""=
></span></p>
</div>
<div>
<p class=3D"x_MsoNormal"><span style=3D""><a href=3D"https://urldefense.com=
/v3/__https://pubs.rsc.org/en/content/ebook/978-1-78262-539-1__;!!PvDODwlR4=
mBZyAb0!X4pZg1VgjczVo7mXiIxE_YflLfRhMwMLn_m8AKHwfgNXyEIl0QeSKpOqjiwxi49ZOE0=
O6mlLZybInA$" originalsrc=3D"https://pubs.rsc.org/en/content/ebook/978-1-78=
262-539-1" shash=3D"uj0NuX/hXgUMcu631m/S1Q8w83Ze25nr+REcXQuys/CldPTZKe3i1u6=
Qn3f4+EQ0y4N8DUVZmfR3AiYLGcesdA4Qb2kDsEMWzoc/QOSgDSqk5WoCbOyvq7KgbOB8DzosRi=
hjELMjxiL50JpLgknM1vZI+MEG522Sm4vjxYrll30=3D"><span style=3D"">PURCHASE
 at https://pubs.rsc.org/en/content/ebook/978-1-78262-539-1</span></a></spa=
n></p>
</div>
<div>
<p class=3D"x_MsoNormal"><span style=3D"">=C2=A0</span></p>
</div>
<div>
<p class=3D"x_MsoNormal"><span style=3D"">University of Toronto<br/>
80 St. George Street<br/>
Toronto, ON<br/>
M5S 3H6<br/>
ph: (416)-978-0448<br/>
fax:=C2=A0 (416)-978-8775</span></p>
</div>
</div>
</div>
</div>
<p class=3D"x_MsoNormal"><span style=3D""><a href=3D"https://urldefense.com=
/v3/__https://sites.chem.utoronto.ca/csicomp/__;!!PvDODwlR4mBZyAb0!X4pZg1Vg=
jczVo7mXiIxE_YflLfRhMwMLn_m8AKHwfgNXyEIl0QeSKpOqjiwxi49ZOE0O6mnfyWo1Lg$" or=
iginalsrc=3D"https://sites.chem.utoronto.ca/csicomp/" shash=3D"LlxhKbiV0OAW=
ksdGegJ3xCWba1g4EciNKxlb3hSdCKjtkZRDYKNIn5/1pbLolojB88F1LlQFQh8thUyIZzRFrdj=
u+oYLee2x4TPcmACfa/RTNlVWP3GtkcL4jO23+hO+mKMctMfHR0FbbnP9nUYjnPlI/jxBsoJJQ2=
yMaXqp+dg=3D"><span style=3D"">https://sites.chem.utoronto.ca/csicomp/</spa=
n></a></span></p>
</div>
</div>
</blockquote>
</div>

</div>




<div width=3D"1" style=3D"color:white;clear:both">_._,_._,_</div>
<hr>


Groups.io Links:<p>


=20=20
    You receive all messages sent to this group.
=20=20
=20=20


<p>
<a target=3D"_blank" href=3D"https://urldefense.com/v3/__https://ammrl.grou=
ps.io/g/main/message/529__;!!PvDODwlR4mBZyAb0!X4pZg1VgjczVo7mXiIxE_YflLfRhM=
wMLn_m8AKHwfgNXyEIl0QeSKpOqjiwxi49ZOE0O6mn_Gc3M7Q$">View/Reply Online (#529=
)</a> |


  <a target=3D"_blank" href=3D"mailto:main_at_ammrl.groups.io?subject=3DRe:%20=
Re%3A%20%5BAMMRL%5D%20upgrade%20to%20TS%202.1%20pl6%20--%3E%20TS%203.6.5%20=
raw%20check%20no%20signal%20on%20gradient%20shimming">Reply To Group</a>
=20=20
    | <a target=3D"_blank" href=3D"mailto:mid15_at_pitt.edu?subject=3DPrivate:=
%20Re:%20Re%3A%20%5BAMMRL%5D%20upgrade%20to%20TS%202.1%20pl6%20--%3E%20TS%2=
03.6.5%20raw%20check%20no%20signal%20on%20gradient%20shimming">Reply To Sen=
der</a>
=20=20

|

  <a target=3D"_blank" href=3D"https://urldefense.com/v3/__https://groups.i=
o/mt/101107231/7559972__;!!PvDODwlR4mBZyAb0!X4pZg1VgjczVo7mXiIxE_YflLfRhMwM=
Ln_m8AKHwfgNXyEIl0QeSKpOqjiwxi49ZOE0O6mnYW_LQtA$">Mute This Topic</a>


| <a href=3D"https://urldefense.com/v3/__https://ammrl.groups.io/g/main/pos=
t__;!!PvDODwlR4mBZyAb0!X4pZg1VgjczVo7mXiIxE_YflLfRhMwMLn_m8AKHwfgNXyEIl0QeS=
KpOqjiwxi49ZOE0O6mkiHbqNrw$">New Topic</a>

<br>




<a href=3D"https://urldefense.com/v3/__https://ammrl.groups.io/g/main/edits=
ub/7559972__;!!PvDODwlR4mBZyAb0!X4pZg1VgjczVo7mXiIxE_YflLfRhMwMLn_m8AKHwfgN=
XyEIl0QeSKpOqjiwxi49ZOE0O6mlYn7vRcQ$">Your Subscription</a> |
<a href=3D"mailto:main+owner_at_ammrl.groups.io">Contact Group Owner</a> |

<div width=3D"1" style=3D"color:white;clear:both">_._,_._,_</div>


--_000_SN4PR04MB83206DF0DEBCA75D5E84FA61DDEBASN4PR04MB8320namp_--

Received on Fri Sep 01 2023 - 18:06:50 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:44:18 MST