--- Hi Robert you need to reset the ACB board. There is a small reset button on the ACB board. Push it it should then reset and work again. --- Hi Robert AN ACB reset will clear the error this can be done either by pressing red reset button on acb board or using acb tool in the utilities. --- Hi Robert, It depends on which vintage you have. There was red reset button (the kind you need a pencil or something sharp to push it with) near where the cord entered the remote keypad if you have one of them. There was a red reset on the ACB which was on the back side of our older console (if I remember right). I think there is a software reset too but not sure. --- Robert: Run a cf on your console. This should clear the Trans P-down light. It will also reinstall all of your configuration parameters so your system should work correctly. --- Maybe the high-power amp for 1h is down and you are only using the low power amp? measuring 90deg pulse length should find if this is the case. --- make sure AQR is on with a pen tip reset ACB execute "ii" command tune and match and retry the experiment --- Robert, Open a terminal window and type acb then type y then type 1 and this should fix the problem. --- Did you do something to turn off the red light next to TRANS PDOWN? We have a DRX300, when we were running xwinnmr, when TRANS PDOWN is on, we could type "acbreset". Now we have TopSpin 1.3, there is no such command anymore, but we do the same by pushing the small "reset" pin on the ACB board by using the tip of a paper clip. Also, "cf" command can achieve the same. --- There are two ways to turn it off: One is double click " bruker utilities" on the desktop, then "servicre tools" then "acb.cmd" then "y", choose "1" to initialize. The light should be off. The second is the push the reset button on acb board with a pen, the light should be off too. Either works. After the light is off, open topspin, then type "ii", you should be back to normal.Received on Fri Sep 04 2015 - 08:43:43 MST
This archive was generated by hypermail 2.4.0 : Mon Jun 19 2023 - 16:35:54 MST