Try to repair redhat system with 'fsck'


> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of daveonofrei via groups.io
> Sent: Tuesday, October 10, 2023 3:57 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Varian communication issue

Hello all,

I'm having an issue with our VNMRS 500 that happened (as per tradition) over
the weekend while I was away. We had a power outage, and while the console
is on UPS and didn't go down, the computer wasn't on backup and shutdown
in the middle of acquisition of a solids Onepul experiment. When I re-booted
the computer, VNMRJ says inactive and won't communicate with the console.
I can ping all of the individual controllers and see data being received
but I cannot get it to go to 'idle'. I'm running VNMRJ 4.2 and so far I've
tried the usual 'su acqproc' to start and stop comms, resetting the master
controller, and power cycling the entire console and computer with no luck.

Any advice would be much very appreciated.

Best,
David O.

-------
David Onofrei, PhD
NMR Facility Manager
Chemistry Department
San Diego State University
work: chem-nmr_at_sdsu.edu<mailto:chem-nmr_at_sdsu.edu>
personal: donofrei2_at_sdsu.edu<mailto:donofrei2_at_sdsu.edu>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#667): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/667__;!!PvDODwlR4mBZyAb0!Qt5mASsLE0TFbeIJ41Zx3BFcXEHGeF02=
SVEunHzyFt5nZTY_MLAwRm5r-qfD8qThOH0aQzv-jat0U0s$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101860941=
/7559972__;!!PvDODwlR4mBZyAb0!Qt5mASsLE0TFbeIJ41Zx3BFcXEHGeF02SVEunHzyFt5nZ=
TY_MLAwRm5r-qfD8qThOH0aQzv-Si9M00c$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 09 2023 - 19:24:55 MST

This archive was generated by hypermail 2.4.0 : Wed Nov 01 2023 - 14:23:34 MST