Re: [AMMRL] Varian communication issue

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Tue, 10 Oct 2023 00:57:31 +0200

Hi David,

have you checked the voltages at the bottom right of the console? All well
a bit above the printed value (more 3.40 than 3.30 V etc.)?
Did you try executing "setacq" from a Linux terminal?

Best regards
Bert

> Am 09.10.2023 um 21:56 schrieb daveonofrei_at_msn.com:
>
> Hello all,
>
> I'm having an issue with our VNMRS 500 that happened (as per tradition) over
> the weekend while I was away. We had a power outage, and while the console
> is on UPS and didn't go down, the computer wasn't on backup and shutdown
> in the middle of acquisition of a solids Onepul experiment. When I re-booted
> the computer, VNMRJ says inactive and won't communicate with the console.
> I can ping all of the individual controllers and see data being received but
> I cannot get it to go to 'idle'. I'm running VNMRJ 4.2 and so far I've tried
> the usual 'su acqproc' to start and stop comms, resetting the master
> controller, and power cycling the entire console and computer with no luck.
>
> Any advice would be much very appreciated.
>
> Best,
> David O.
>
> -------
> David Onofrei, PhD
> NMR Facility Manager
> Chemistry Department
> San Diego State University
> work: chem-nmr_at_sdsu.edu
> personal: donofrei2_at_sdsu.edu
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software http://www.mestrelab.com
Barthel HF Amplifiers http://www.barthel-hf.de


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#668): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/668__;!!PvDODwlR4mBZyAb0!Q9DmZZdcFk1HTKowtMQfa3bSEVotqUr4=
0mhxvu35k4bXrKDhUG3hf5D-NoNBFOX9YRcPM1a6uqU0gPE_5sU6rg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101860941=
/7559972__;!!PvDODwlR4mBZyAb0!Q9DmZZdcFk1HTKowtMQfa3bSEVotqUr40mhxvu35k4bXr=
KDhUG3hf5D-NoNBFOX9YRcPM1a6uqU0gPGS6C1hFA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 09 2023 - 19:25:14 MST

This archive was generated by hypermail 2.4.0 : Wed Nov 01 2023 - 14:23:34 MST