Re: diffusion of J-coupled spins

Sambasivam Sendhil Velan (velan@gg.caltech.edu)
Tue, 21 Dec 1999 17:01:22 -0800 (PST)

Measuring the self - diffusion of J coupled spins.

Dear Dennis,

The following approaches are possible:

In the standard PGSE / PGSTE experiments you can keep the capital delta
fixed (the diffusion encoding period and increment the small delta) so
that the J modulation remains the same for all the gradient strengths.

There are other approaches to measure self diffusion of coupled spins

S. Sendhil Velan and N. Chandrakumar, JMR Series A, 122 - 125, 1996.

Use the chemical shift selective approach and this will enable you to
measure the self diffusion of individual components in multi-component
system. This method looks at the line widths which is a measure of
frequency as compared to other methods which measures the amplitude.

Sendhil Velan

*******************************************************************************
Sambasivam Sendhil Velan