Hello Mike,
If you have an additional Gradient filter, please consider changing it to see
if that can resolve the problem.
I have attached a picture of one of those Gradient filters for reference.
https://drive.google.com/file/d/1JEhiqXDmFEnscm0cYo2jZRG8yOVLVhIQ/view?usp=share_link
Regards
Vas
On Thu, May 18, 2023 at 02:16 PM, Michael Groves wrote:
>
>
>
> spicious that you get a GCU error and then when you kill the acquisition
> they calm down. It might be worth disconnecting the gradient cable from
> the probe the next time you see them to see if it's something strange
> happening with the gradient amplifier/system/something. If you're getting
> RCU and GCU errors, it also might be worth considering the CCU -- that's the
> brains of the AQS rack and if it’s losing its mind then it could cause
> issues for multiple boards downstream.
>
>
>
>
>
>
>
> Based on the acquisition errors you're seeing I don't think it's anything
> in the BSMS. Right now I think the lock issues are a symptom and not the
> cause of the problem.
>
>
>
>
>
>
-=-=-=-=-=-=-=-=-=-=-=-