We've had similar experiences with Topspin 3.5pl2 and 3.5pl4, but not
earlier versions.

--Mike


-----------------------------
Michael Strain, PhD
Director, CAMCOR NMR Facility
University of Oregon
Eugene, OR 97403-1253
mstrain_at_uoregon.edu
541-346-4605 office/lab
541-556-4077 mobile
http://nmr.uoregon.edu





On Tue, 3 May 2016, Paul White wrote:

> Hi John,
>
> Thanks for bringing this up! I am recently having the exact same problem as
> well. It locks out most of the digital BSMS and to reset things back to
> normal I hard restart the BSMS/2 followed by "cf". For some reason it also
> prevents a soft BSMS shutdown from succeeding, gives the error the ELCB is
> unresponsive and it loses the heartbeat for the BSMS.
>
> It also happens whenever I try to use new solvents that I create using edsolv
> or edlock...even when the new solvents are just copy/paste of existing
> generic solvents but renamed!
>
> Are you using TopSpin 3.5 pl 5? I have not had this happen with TopSpin 3.0
> or TopSpin 3.2 but when we upgraded it started happening with MeOH-d4. Maybe
> it's a firmware issue?
>
> I would love to hear the responses you receive to this.
>
> Sincerely,
> Paul
>
> Dr. Paul White
> Institute for Molecules and Materials
> Radboud University Nijmegen
> +31 (0)24 365 2959
> P.White_at_science.ru.nl
>
>
>
>
>
>
>
> On 5/3/16 12:28 AM, John N Glushka wrote:
>> I’m new to Bruker, and just had an error trying to use topshim with MeOD-4.
>> It refers to the use of the shaped pulse in the topshim1d pulseprogram used
>> for methanol and states that SP0=0.0 and is too short and should be at
>> least 0.25us. We have used methanol in the recent past without a problem,
>> and so I don’t understand why this parameter is now causing problems, and
>> how one can fix it. Possibly related to this was an earlier error during
>> locking referring to BSMS control; lock key wasn’t valid. I reset the ELCB
>> which seemed to clear that up. Any suggestions would be welcome-
>> thanks,
>> John
>>
>>
>>