Re: AMMRL: topshim1d SP0 question

From: Paul White <P.White_at_science.ru.nl>
Date: Tue, 3 May 2016 21:08:16 +0200

Hi John,

Thanks for bringing this up! I am recently having the exact same problem
as well. It locks out most of the digital BSMS and to reset things back
to normal I hard restart the BSMS/2 followed by "cf". For some reason it
also prevents a soft BSMS shutdown from succeeding, gives the error the
ELCB is unresponsive and it loses the heartbeat for the BSMS.

It also happens whenever I try to use new solvents that I create using
edsolv or edlock...even when the new solvents are just copy/paste of
existing generic solvents but renamed!

Are you using TopSpin 3.5 pl 5? I have not had this happen with TopSpin
3.0 or TopSpin 3.2 but when we upgraded it started happening with
MeOH-d4. Maybe it's a firmware issue?

I would love to hear the responses you receive to this.

Sincerely,
Paul

Dr. Paul White
Institute for Molecules and Materials
Radboud University Nijmegen
+31 (0)24 365 2959
P.White_at_science.ru.nl







On 5/3/16 12:28 AM, John N Glushka wrote:
> I’m new to Bruker, and just had an error trying to use topshim with MeOD-4. It refers to the use of the shaped pulse in the topshim1d pulseprogram used for methanol and states that SP0=0.0 and is too short and should be at least 0.25us. We have used methanol in the recent past without a problem, and so I don’t understand why this parameter is now causing problems, and how one can fix it. Possibly related to this was an earlier error during locking referring to BSMS control; lock key wasn’t valid. I reset the ELCB which seemed to clear that up. Any suggestions would be welcome-
> thanks,
> John
>
>
>
> John Glushka
> Complex Carbohydrate Research Center
> University of Georgia
> 315 Riverbed Rd.
> Athens, GA 30602
> glushka_at_uga.edu
> 706-542-4483
>
>
>
Received on Tue May 03 2016 - 09:07:52 MST

This archive was generated by hypermail 2.4.0 : Mon Jun 19 2023 - 17:44:28 MST