Re: AMMRL: topshim1d SP0 question

From: Liermann, Dr. Johannes <liermann_at_uni-mainz.de>
Date: Wed, 4 May 2016 18:44:39 +0000

Dear colleagues,


what you describe sounds like it could be related to Prosol which had the odd habit of losing some pulse durations during the transition from TopSpin 3.2 to 3.5 at our workstations.

Please type „edprosol“ and check if there are 2H observe and decouple 90° pulses for the solvents affected. If not, copy some reasonable values from another solvent and have Prosol recalculate your pulse set for that solvent.

I hope that helps.


Best regards,
Johannes Liermann


---------------------------------------------------------------
Dr. rer. nat. Johannes Liermann
Johannes Gutenberg-Universität Mainz
Institut für Organische Chemie
Duesbergweg 10-14
55128 Mainz

Tel. (06131) 39-25466
Fax (06131) 39-25454
liermann_at_uni-mainz.de

http://www.nmr.oc.chemie.uni-mainz.de/










Am 03.05.16, 21:08 schrieb "Paul White" <P.White_at_science.ru.nl>:

>Hi John,
>
>Thanks for bringing this up! I am recently having the exact same problem
>as well. It locks out most of the digital BSMS and to reset things back
>to normal I hard restart the BSMS/2 followed by "cf". For some reason it
>also prevents a soft BSMS shutdown from succeeding, gives the error the
>ELCB is unresponsive and it loses the heartbeat for the BSMS.
>
>It also happens whenever I try to use new solvents that I create using
>edsolv or edlock...even when the new solvents are just copy/paste of
>existing generic solvents but renamed!
>
>Are you using TopSpin 3.5 pl 5? I have not had this happen with TopSpin
>3.0 or TopSpin 3.2 but when we upgraded it started happening with
>MeOH-d4. Maybe it's a firmware issue?
>
>I would love to hear the responses you receive to this.
>
>Sincerely,
>Paul
>
>Dr. Paul White
>Institute for Molecules and Materials
>Radboud University Nijmegen
>+31 (0)24 365 2959
>P.White_at_science.ru.nl
>
>
>
>
>
>
>
>On 5/3/16 12:28 AM, John N Glushka wrote:
>> I’m new to Bruker, and just had an error trying to use topshim with MeOD-4. It refers to the use of the shaped pulse in the topshim1d pulseprogram used for methanol and states that SP0=0.0 and is too short and should be at least 0.25us. We have used methanol in the recent past without a problem, and so I don’t understand why this parameter is now causing problems, and how one can fix it. Possibly related to this was an earlier error during locking referring to BSMS control; lock key wasn’t valid. I reset the ELCB which seemed to clear that up. Any suggestions would be welcome-
>> thanks,
>> John
>>
>>
>>
>> John Glushka
>> Complex Carbohydrate Research Center
>> University of Georgia
>> 315 Riverbed Rd.
>> Athens, GA 30602
>> glushka_at_uga.edu
>> 706-542-4483
>>
>>
>>
>
Received on Wed May 04 2016 - 08:44:15 MST

This archive was generated by hypermail 2.4.0 : Mon Jun 19 2023 - 17:44:28 MST