Re: [AMMRL] NEO 600 error message- No topshim

From: Clemens Anklin via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Fri, 30 Jun 2023 15:00:04 +0000

Dear All

running the installnmr script will set the permission of the files in the topspin
folder more nuanced than a chmod. The script is found in <topspinhome>/prog/bin

Clemens


On 6/30/2023 7:56 AM, ssham_at_luc.edu<mailto:ssham_at_luc.edu> wrote:
**EXTERNAL EMAIL**
Hi Martha,
I had the same Topshim permission problem on Centos 7 with Topspin 4.
something, but not the second part of your problem.
First, I tried to change the permissions of the topshim folder (a+xwr). It
did not work.
Finally, I did the same from the top of the parent directory. And it worked.
Don't ask me why!

Good luck.

Simon



--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 twitter: _at_canklin
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com<mailto:clemens.anklin_at_bruker.com>
==========================
==========================
====================


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#270): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/270__;!!PvDODwlR4mBZyAb0!UGwxi-26gg1MeTRMrexZAXVr4xZpkEKO=
B2rquYmHO1C4Hfd_2_1WXPaSrjTlmPtSH0m1ms_Wyhe6G2KgKYHKj2bqSA4T$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99865246/=
7559972__;!!PvDODwlR4mBZyAb0!UGwxi-26gg1MeTRMrexZAXVr4xZpkEKOB2rquYmHO1C4Hf=
d_2_1WXPaSrjTlmPtSH0m1ms_Wyhe6G2KgKYHKj90dUYHF$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 30 2023 - 08:00:46 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST