Your problem seems bigger than this solution, but generally, when I run
into weird permissions issues, I run the $TSHOME/prog/bin/installnmr
program (the same one you'd use when you change the NMR admin password). It
resets all the permissions inside the Topspin directory to their defaults.
On Thu, Jun 29, 2023, 8:37 PM Martha Morton <mmorton4_at_unl.edu> wrote:
> Dear group,
>
> I'm running Topspin 4.1.3 on a 600 NEO console. The operating system is
> CentOS 7. Every time I update the distro topshim breaks. Initially, I saw
> errors in permissions to use shim files. Now I'm seeing errors where
> topshim will start then stop throwing the error "setupHardpulseP1plw1 - F=
or
> logic channel 1 power level not determined ".
> Has anyone seen and solved this?
>
> Thanks,
> Martha
>
> Sent from my Verizon, Samsung Galaxy smartphone
>
>
-=-=-=-=-=-=-=-=-=-=-=-