Re: [AMMRL] Unexpected peaks of 19F NMR from CF3CH2OH

From: Kirk Marat <kirk.marat_at_gmail.com>
Date: Wed, 12 Jul 2023 11:29:11 -0700

There is a significant 13C-19F isotope shift.

Kirk Marat
Victoria, BC

On Wed., Jul. 12, 2023, 11:22 Holmes, Daniel, <holmesd5_at_msu.edu> wrote:

> The 13C satellites are not symmetric about the 19F-12C peak due to the
> heavy isotope effect. See Glen Facey’s excellent post about it.
> http://u-of-o-nmr-facility.blogspot.com/2012/10/isotope-effects-and-19-f-13-c-hmqc.html
>
> Dan
>
>
> *From: *main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Weidong
> Hu via groups.io <whu=coh.org_at_groups.io>
> *Date: *Wednesday, July 12, 2023 at 1:54 PM
> *To: *main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> *Subject: *[AMMRL] Unexpected peaks of 19F NMR from CF3CH2OH
>
> Hi All,
>
> I acquired 1D-19F NMR of CF3CH2OH dissolved in CDCl3/CD3OD(2:1 v/v) on CP
> TCI 700 H&F-C/N-D probe (without 1H decoupling), and the spectrum is very
> confusing.
>
> In addition to main triplet peaks, there are three minor triplet peaks
> (about 0.5% of main triplet). The positions of minor peaks are not
> symmetric to the main triplet, and thus should not come from natural
> abundance of 13C-19F J coupling. It is reasonable to see four minor peaks
> generated from one and two-bond J couplings between 13C-19F, but they
> should be symmetric to the main peak. I don't know how to explain the
> observed spectrum. Any suggestions as what is going on? The spectrum is
> attached below.
>
> Thank you,
>
> Weidong
>
> Dr. Weidong Hu
> Department of Immunology & Theranostics
> Beckman Research Institute
> 1450 E. Duarte Rd., Duarte, CA 91010
> Tel: 626 256 4673, Ext: 63416
> Fax: 626 301 8186
>
>
> ------------------------------
>

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#307): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/307__;!!PvDODwlR4mBZyAb0!TL2ZjYgaHyEFpyQOVitUw6mqgZUOVPvZ=
DuVsLzeVxPYkVGXZwABG-bYzoOk5hhoCy_GfQY4jh_1-vNPduw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100104757=
/7559972__;!!PvDODwlR4mBZyAb0!TL2ZjYgaHyEFpyQOVitUw6mqgZUOVPvZDuVsLzeVxPYkV=
GXZwABG-bYzoOk5hhoCy_GfQY4jh_0WlVDhXg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jul 12 2023 - 12:01:49 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:02 MST