[AMMRL] Unexpected peaks of 19F NMR from CF3CH2OH

From: Weidong Hu via groups.io <whu=coh.org_at_groups.io>
Date: Wed, 12 Jul 2023 17:33:28 +0000

Hi All,

I acquired 1D-19F NMR of CF3CH2OH dissolved in CDCl3/CD3OD(2:1 v/v) on CP TCI
700 H&F-C/N-D probe (without 1H decoupling), and the spectrum is very confusing.

In addition to main triplet peaks, there are three minor triplet peaks (about
0.5% of main triplet). The positions of minor peaks are not symmetric to the
main triplet, and thus should not come from natural abundance of 13C-19F J coupling.
It is reasonable to see four minor peaks generated from one and two-bond J couplings
between 13C-19F, but they should be symmetric to the main peak. I don't know how to
explain the observed spectrum. Any suggestions as what is going on? The spectrum is
attached below.

Thank you,

Weidong

Dr. Weidong Hu
Department of Immunology & Theranostics
Beckman Research Institute
1450 E. Duarte Rd., Duarte, CA 91010
Tel: 626 256 4673, Ext: 63416
Fax: 626 301 8186

[cid:image004.png_at_01D9B4AC.4AC83670]




----------------------------------------------------------------------
------------------------------------------------------------
-SECURITY/CONFIDENTIALITY WARNING-

This message and any attachments are intended solely for the individual or
entity to which they are addressed. This communication may contain information
that is privileged, confidential, or exempt from disclosure under applicable
law (e.g., personal health information, research data, financial information).
Because this e-mail has been sent without encryption, individuals other than
the intended recipient may be able to view the information, forward it to others
or tamper with the information without the knowledge or consent of the sender.
If you are not the intended recipient, or the employee or person responsible for
delivering the message to the intended recipient, any dissemination, distribution
or copying of the communication is strictly prohibited. If you received the
communication in error, please notify the sender immediately by replying to this
message and deleting the message and any accompanying files from your system.
If, due to the security risks, you do not wish to receive further communications
via e-mail, please reply to this message and inform the sender that you do not
wish to receive further e-mail from the sender. (LCP301)
------------------------------------------------------------


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#302): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/302__;!!PvDODwlR4mBZyAb0!RV4Tm975zvGwv_cpbI_UcrYB_hjqQD_S=
r4rvARbgfpk1xNqXdThqhKEXiPzXl1-AzlDZYoCvAzlUO3lFSilhtg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100104757=
/7559972__;!!PvDODwlR4mBZyAb0!RV4Tm975zvGwv_cpbI_UcrYB_hjqQD_Sr4rvARbgfpk1x=
NqXdThqhKEXiPzXl1-AzlDZYoCvAzlUO3l9NBHYig$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






image004.png
(image/png attachment: image004.png)

Received on Wed Jul 12 2023 - 10:54:04 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:02 MST