Hi Weidong,
It is the C13 isotope effect on the F19 chemical shifts. The chemical shift
of the F19 bound to the C13 is in the middle of the peak 1 and peak 3. The
chemical shift of the F19 two bonds away from C13 is in the middle of peak
2 and peak 4 which overlaps the main peak.
Best,
Ion
Ion Ghiviriga
Director of the Center for NMR Spectroscopy
University of Florida, Chemistry Department
Gainesville, Fl 32611-7200
Ph: 352-846-3001
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Weidong Hu v= ia groups.io
Sent: Wednesday, July 12, 2023 1:33 PM
To: main_at_ammrl.groups.io
Subject: [AMMRL] Unexpected peaks of 19F NMR from CF3CH2OH
Hi All,
I acquired 1D-19F NMR of CF3CH2OH dissolved in CDCl3/CD3OD(2:1 v/v) on CP TCI
700 H&F-C/N-D probe (without 1H decoupling), and the spectrum is very confusing.
In addition to main triplet peaks, there are three minor triplet peaks (about
0.5% of main triplet). The positions of minor peaks are not symmetric to the
main triplet, and thus should not come from natural abundance of 13C-19F J
coupling. It is reasonable to see four minor peaks generated from one and
two-bond J couplings between 13C-19F, but they should be symmetric to the
main peak. I don't know how to explain the observed spectrum. Any suggestions
as what is going on? The spectrum is attached below.
Thank you,
Weidong
Dr. Weidong Hu
Department of Immunology & Theranostics
Beckman Research Institute
1450 E. Duarte Rd., Duarte, CA 91010
Tel: 626 256 4673, Ext: 63416
Fax: 626 301 8186
________________________________
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#305): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/305__;!!PvDODwlR4mBZyAb0!Tv6PPbmZKbvW0Wig73vGmS09GTi_3gbs=
7A0oYXXNn6pJeDwtRaUhycNTGPxRFxmzhyM9EA-8Vdj_xAY$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100104757=
/7559972__;!!PvDODwlR4mBZyAb0!Tv6PPbmZKbvW0Wig73vGmS09GTi_3gbs7A0oYXXNn6pJe=
DwtRaUhycNTGPxRFxmzhyM9EA-8oIM6xfI$
Group Owner: main+owner_at_ammrl.groups.io
Received on Wed Jul 12 2023 - 12:01:51 MST