Re: How to remove helium gas in case of a quench

From: Ben Bangerter <ben.bangerter_at_yale.edu>
Date: Thu, 20 Dec 2001 12:10:47 -0500

Yes, you will need a quench vent, piped to the outside.. The Oxford 800 here
has a big flange (about 6" inside diameter) on the dewar, and this is
connected to a run of ~8" plastic pipe to a hole in the wall to the outside.
When this magnet quenches, it is a sight to behold (and hear). The dogwood
tree about 15' away from the vent outside experienced an "early autumn" when a
bunch of its foliage got quick-frozen by the first of two (or three, I forget)
quenches experienced during charging about 4 years ago. This magnet is all by
itself in a room about 18' square.

Ben Bangerter

********************

"Zhang, Weixing" wrote:

> Hello all:
>
> We are planning to install a shielded 800 MHz magnet in a relatively small
> room
> (30 x 40 ft, 8 ft ceiling height). The magnet is going to sit in a dip.
> In case of a quench, the amount of helium gas will be huge for a shielded
> 800 magnet. Has anyone installed such magnet in a relatively small room?
> What methods (fan or pipe to the outside) you planned to use to remove the
> helium gas in case of a quench?
>
> Any comment is appreciated.
>