Re: [AMMRL] pipes for helium gas

From: Ulrich <Ulrich.Haunz_at_uni-konstanz.de>
Date: Thu, 17 Aug 2023 13:59:23 +0200

Dear Dr. Tošner!

Am Donnerstag, August 17, 2023 11:34 CEST, schrieb Zdeněk Tošner <zdenek.tosner_at_natur.cuni.cz>:

> Dear AMMRL,
> we are thinking of collecting helium boil-off from our four magnets located
> in two distinct rooms. I would be grateful for the experience how to bring
> helium gas to a small (500l) collection bag which will be located 40m away
> from magnets. Boil-off rate is rather small and so is the helium gas flow
> (about 30 liters of gas per hour, if I estimate correctly). Are the steel
> pipes necessary? Can we use some other material like PE or PTFE tubes? And
> does the diameter of pipes play a role for such small flows?

Helium is a bit tricky - please use Copper pipelines - there is a special
Copper alloy and treatment for the Helium lines. PE and PTFE are not Helium
tight enough!

We have a 'backbone' on these Copper lines through our buildings. We are
attached to the building of the Physic department. And these have a recycling
and liquification machines in the cellar. On the top of the building is the
Helium balloon, filling this 10 m3 vessel up. When the balloon=
is quite full, then the Helium gas is compressed into cylinders and stored
there, until it is liquidified to liquid Helium. The team of the next department
controls the gas, if there is a leak, it will be immediately closed.=
The problem is, the liquification process is cleaning the gas before it becomes
liquid - and this needs a lot of time and energy, to get rid of oxygen, nitrogen
and vapour of water. If the gas is quite pure, the rate of the machine gets
higher.

Bruker delivers a small station with a small balloon - if the balloon is full,
then the gas is compressed into a set of cylinders. And the platform with
the cylinders can be sent to a liquification process. Then the liquified
Helium is then delivered to you.

Heliosmart is the brand name by Bruker - at the end of the magnet is the
Heliosmart parts:

https://www.bruker.com/en/products-and-solutions/mr/nmr/ascend-nmr-magnets.html

You can also can do it with your own system.

Yours sincerely,

Ulrich Haunz

> Best regards,
> Zdeněk Tošner
> --
>
> Laboratory of NMR Spectroscopy
> Charles University in Prague
> Faculty of Science
> Albertov 6, 128 43 Praha 2
> http://www.natur.cuni.cz/en
>
>
>
>
>



--
=========================

Ulrich Haunz, Dipl.-Ing. (FH)
NMR-Spektroskopie

Raum L 710, L 807
NMR-Center M520, M521, M527
Postfach M 715
Fachbereich Chemie
Universitaet Konstanz

78457 Konstanz

Tel.: +49 (0)7531 88-4261 -> M521 Bruker Avance NEO 800
Tel.: +49 (0)7531 88-4422 -> L807 JEOL Lambda 400
Tel.: +49 (0)7531 88-4955 -> L710 Bruker AVIII HD 400
Tel.: +49 (0)7531 88-4956 -> M520a Bruker AVIII 600
Tel.: +49 (0)7531 88-4957 -> M527 Bruker AVIII 400 Automation
Tel.: +49 (0)7531 88-4958 -> M520 Bruker AVIII 400 WB
Tel.: +49 (0)170 9153213
FAX : +49 (0)7531 88-3898
=========================


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#457): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/457__;!!PvDODwlR4mBZyAb0!U4V619W-_dyU-Pkc_lW5o35ka0JZhXn9=
8xXPq__P1EYsDOnxfVwzu-g5a1eccYLLS0ATSx3TZ-I98NI4Q0RmmOZvhBriFA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100797790=
/7559972__;!!PvDODwlR4mBZyAb0!U4V619W-_dyU-Pkc_lW5o35ka0JZhXn98xXPq__P1EYsD=
OnxfVwzu-g5a1eccYLLS0ATSx3TZ-I98NI4Q0RmmOYIi_xH6w$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Thu Aug 17 2023 - 05:10:59 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:56 MST