[AMMRL] Pipes of helium - Heliumleitungen

From: Ulrich <Ulrich.Haunz_at_uni-konstanz.de>
Date: Thu, 17 Aug 2023 14:29:26 +0200

Dear Zdenek,

there are two aspects.

  1. You loose about 30% of helium gas during the refill. Interesting question,
why. Nevertheless it's a rather well known value, even more, I measured multiple
times during my active time.

To tell you a consequence. For our 900 MHz Magnet we had a pipeline with an
inner diameter of 25 mm with a length Of about 50 meter until the next bigger
pipeline. The result was a filling time of 4 hours if we accept an overpressure
of 50 mBar during the filling procedure. That's a lot. Its not dangerous, but
increases the temperature of the helium by a few millikelvin. If you release
the overpressure immediately, this results in evaporating liquid helium from
100% to 93%, which is alot again. We extended the filling time to 5 hours and
released the overpressure slowly.
Everything is possible, but slow.

  2. If you use any plastics material except HD-PE you get significant diffusion
of oxygen and nitrogen inside the tube. Depends on your liquefier, how much you
can accept.

Have a nice day

Rainer Haessner



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#459): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/459__;!!PvDODwlR4mBZyAb0!S32GmN1FpWghpnNNocLySvcjGiggIxT6=
swIhe63QMcWC70Z2xQFWnpWgfXbEHZ0TE0CbnEjlfyMCihHDT5QYl4Qm35U2yw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100801085=
/7559972__;!!PvDODwlR4mBZyAb0!S32GmN1FpWghpnNNocLySvcjGiggIxT6swIhe63QMcWC7=
0Z2xQFWnpWgfXbEHZ0TE0CbnEjlfyMCihHDT5QYl4RfRx8eBw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Thu Aug 17 2023 - 07:16:59 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:56 MST