Re: [AMMRL] Cortab problem after updating to TS 3.8.0 ?

From: Harry Mackenzie via groups.io <harry.mackenzie=chem.ox.ac.uk_at_groups.io>
Date: Thu, 6 Aug 2026 13:59:16 +0000

Thanks for all the advice everyone – it indeed turns out that I had
done the expinstall command too early (ie before cf) and so repeating the
process sorted it all out 😊

Thanks again, Harry

> From: main_at_ammrl.groups.io On Behalf Of Marsden, Ian via groups.io
> Sent: 05 August 2026 2:52 pm
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Cortab problem after updating to TS 3.8.0 ?

Harry:

    One thing that I might suggest if you didn't
do it is to run "expinstall" after setting things in cf. There is usually a reminder
to do that, so I assume you did, but we sometimes see these issues when we change routings.

Good luck,

Ian.

Ian Marsden, PhD

Principal Research Scientist I
Small Molecule Analytical Sciences, AP9 Space ID LL014A
1 N. Waukegan Road
North Chicago, IL 60064-6114
USA

OFFICE +1 847-935-4604
EMAIL ian.marsden_at_abbvie.com



http://www.abbvie.com/

________________________________
> From: main_at_ammrl.groups.io on behalf of Harry Mackenzie via groups.io
> Sent: Wednesday, August 5, 2026 4:34 AM
> To: main_at_ammrl.groups.io
> Subject: [EXTERNAL] [AMMRL] Cortab problem after updating to TS 3.8.0 ?


Hi folks,

Hoping someone can help me with cortab…

I had to replace the controller PC and have installed Topspin 3.8.0 without
issue.

I restored the previous installation files using nmr_save and everything appeared rosy.

However, when trying to shim on the sucrose sample (using 1H), I get this error:

and then this one:

Seems to be an issue with cortab yet the green ‘cortab available’
light shows in edasp and one can acquire 1H data without issue:

There are a bunch of files in the <TSHOME>/conf/instr/<spec>/cortab/ folder
but the cortab window looks a bit sparse to me:

Can anyone point me in the right direction please?

Many thanks,

Harry

Dr. Harold W. Mackenzie MRSC
Head of Organic and Biological NMR
a: Department of Chemistry, Mansfield Road, Oxford, OX1 3TA
t: +44 (0) 1865 275620
e: harry.mackenzie_at_chem.ox.ac.uk
w: nmrchem.web.ox.ac.uk


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3668): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3668__;!!PvDODwlR4mBZyAb0!WtFblYAtA_p8IkZpnj3JJkaLCSc91I=
_VR7H9Of5_DQ2VPS3oB2Gs7YWmiudY64Vzhkw9v8VU0Ez2ifX9i2J4ETtXrGh7$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120607753=
/7559972__;!!PvDODwlR4mBZyAb0!WtFblYAtA_p8IkZpnj3JJkaLCSc91I_VR7H9Of5_DQ2VP=
S3oB2Gs7YWmiudY64Vzhkw9v8VU0Ez2ifX9i2J4EcJ-ar14$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Aug 06 2026 - 08:49:23 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 15 2026 - 14:28:24 MST