I have no authoritative information, but I have had experience with
running computers and backup devices near magnets. Our Mac IIfx is
well within 10G line of our 300 MHz wide bore, and the computer itself
and hard disk have given no problems. We have had a rewritable
optical drive fail, but its replacement has proved much more durable,
and I don't attribute the failure to the field. The floppy does seem
to have problems, and presently is not working at all, but who knows
why floppy drives fail. Some of our Sun boxes are in >10G field, and
the only failure we have seen was in a DAT tape drive (WangDAT) that
kept eating tapes. It continued to do so long after it was moved away
from the field, so I cannot be certain the field had any effect, but
we have kept our new one (an HP, with a different tape handling
mechanism) farther away from the magnets and it has given no problems.
The general consensus seems to be that you need to keep the general
population outside the 5 G line, and that warnings must be posted and
traffic limited inside it. In your situation the people directly
below will probably be working close to the 5G line, and you may have
to have someone make measurements if the room is to be general access.
They will certainly see distortions in their computer monitors, and
may need to be made aware of field effects. I usually don't worry
about having a floppy disk or a magnetic tape in a 5G field, but you
have to be careful of other things. Steel tools and other objects
become magnetized. The local fields from these objects may be much
greater than the steady stray field that magnetized them, so you can
create a whole army of harmless-looking things that can damage data if
floppy disks or tapes are set next to them. I have never seen
warnings about this, and would be happy to learn that my fears are
groundless, if someone has better information. That is why I am
posting this back to the ammrl too.
-- Steve Philson philson@nmr.chem.umn.edu Director NMR Lab 612-626-0297 Chemistry Dept. University of Minnesota