At 13:07 2/21/95 -0800, Michael Strain wrote:
>On Tue, 21 Feb 1995, Marty Pagel wrote:
>
>> AMMRL,
>>
>> Our house nitrogen is being shut off permanently by the end of this month.
>> We plan to use the boil-off from two 5-foot-tall N2 dewars which we'll plumb
>> into the old house N2 lines in our lab facility. Has anyone else had
>> experience with setting up an isolated, stable nitrogen gas delivery system
>> in your NMR lab/facility?
>>
>
>For routine VT/eject/spin we use compressed air from a Balston type 75-20
>air dryer/purifier. A single unit is used to supply four NMR
>spectrometers with "-100 deg F" dewpoint air. The filters also remove oil
>and particulates containing nasty metals and their oxides, like iron &
>copper. It cost about $2500 in 1992, and has functioned flawlessly. We
>estimate that the payback time was barely a year, compared with using
>liquid N2 boil-off.

For those who don't have sufficient house pressure to operate the 75-20 (60
psig is required, which we don't have), Balston also makes a complete unit
with compressor (74-5041 - sold as an FTIR purge gas generator). The
capacity is quite a bit lower (one NMR machine, max), and the price quite a
bit higher, but it is definitely a possibility for a completely independent
facility. We have one here, and I'll be glad to share experiences with
anyone who is interested.