We also use the Balston air drier/filter systems for all the NMR
spectrometers. However, should you want to use variable temperature with a
liquid nitrogen bath you will need to switch over to nitrogen gas for
spinning, etc. Therefore we have a nitrogen manifold in addition to the air
manifold. There are labeled manual valves on the wall at each spectrometer
to accomplish the switching. We purchase 160 liter tanks of liquid nitrogen
with a heat exchanger that delivers room temperature nitrogen at approx. 200
psi. (I have heard these variously referred to as "VGL's" or "gas packs".
I am sure your liquid nitrogen supplier would know what they are. We use 2
of these attached to an Air Products "LC-91 Liquid Gas Manifold System"
which will automatically switch to the second dewar when the pressure in the
first drops too low.
Jane Strouse
Dept. of Chemistry
and Biochemistry
UCLA