FTS Chiller

Karen Ann Smith (karenann@unm.edu)
Tue, 21 Aug 2001 12:51:55 -0600


I have a FTS XR1185 chiller.   I am using it to keep a peptide at around
6C so I don't need it to go too low, but I do need it to stay cold for a
long time.

I have been using house nitrogen (liquid N2 boiloff) after my dried air
froze up the chiller a couple of week ago.  Now I purge the lines for a
good 10 minutes before starting the chiller.

I can get the temperature at the chiller down to -66 or so.  It stays
there for at least 6 hours (probably 12-18) and then starts to warm up. 
This morning when I came in, the chiller was at -45 and the sample at
about +15. 

There did not seem to be any blockage of the air flow.  I was running at
about 535liter/hour which is quite a bit less than the 1680 liter/hour
(1scf/m) that the system is rated for.

The room temperature was at a reasonable 72F this morning, so room
temperature was probably not a factor.

So- should I try running at 1600 liters/hour?  Won't that bounce the
sample around?

Is there some other trick, or should I just revert to liquid nitrogen
boiloff?  (How long does a Bruker 25L dewar last, anyway?)