Hello Ann,
You could type the command: iexpno
As the name implies, it "increments" the experiment number, while taking
all of the parameters from the current dataset.
If the next dataset already exists, then you simply move to the new
dataset (nothing is overwritten). But if it doesn't exist, then a new
dataset is created with all of the current parameters but no data.
I am too lazy to type "iexpno" so I wrote a one line macro called: i
The macro calls "iexpno"
Cheers,
Bob.
On 2023-09-14 1:05 p.m., Anne Carroll via groups.io wrote:
>
>
>
> Hello,
> Is there a way to create a new EXPNO from the TopSpin command line,
> avoiding the interactive NEW experiment window?
> I would like to be able to create a new EXPNO in the current dataset
> with the current parameters as part of a chain of TopSpin commands.
>
> Thank you,
> Anne Carroll
> --
> Anne M. Carroll, Ph.D. (she/her) NMR/MS Instrumentation Specialist
> Department of Chemistry and Biochemistry Miami University Oxford, Ohio
> USA513-529-3163 Office 678-637-6739 Cell carrol50_at_miamioh.edu
>
--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.
McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With OneSpoon" wampum agreement.
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#578): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/578__;!!PvDODwlR4mBZyAb0!Um561Dmak_PzhE47f9C4OTAKYNJLXUTq=
MlcfUTVdc38s2BhPXERkBjyfoFdKtpXUKbRs700WccW-wXuD$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101363004=
/7559972__;!!PvDODwlR4mBZyAb0!Um561Dmak_PzhE47f9C4OTAKYNJLXUTqMlcfUTVdc38s2=
BhPXERkBjyfoFdKtpXUKbRs700WceVLgIhf$
Mute #bruker:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!Um561Dmak_PzhE47f9C4OTAKYNJLXUTqMlcfUT=
Vdc38s2BhPXERkBjyfoFdKtpXUKbRs700WcUWFqd1j$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Sep 14 2023 - 10:40:49 MST