Re: [AMMRL] Oscillating wave on WOBB curve

From: johnhahn cazent.com <johnhahn_at_cazent.com>
Date: Thu, 14 Sep 2023 17:16:42 +0000

Hi Sameer,

I think Jon is on the right path....if this instrument has a switching power
supply, the ripple won't be 60Hz however...it will be in the vicinity of
140-180KHz depending on the load and design of the tank...do you have the
wherewithal to check the AC noise ripple on this instrument?