Re: [AMMRL] 2.5mm spin problems

From: Garimella, Ravi <rgarimel_at_odu.edu>
Date: Wed, 6 Sep 2023 18:42:33 +0000

MAS2 configured fine, 2.5mm probe selected and configured fine

Checked Manually changing BP/DP when the cover is opened and holding it up
- does not affect or show any spinning and it fails

4mm spins fine


> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Michael Groves via groups.io
> Sent: Wednesday, September 6, 2023 2:32 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] 2.5mm spin problems


 
Is the MAS/2 correctly configured-2.5mm probe selected? And is the 2.5mm probe
correctly configured? If it's a flip-type, does the MAS controller have that
box checked? If you disconnect the bearing and drive air lines from the probe
and manually open the valves in the MAS/2 (use the manual spin control settings
on the touchscreen) does the air turn on and ramp up and down as expected? Your
website says you've got a couple of 4mm probes-do they spin okay? When you try
to start the sample spinning, can you hear it spinning-do you hear the whine
ramping up, or do you not hear anything?

Cheers,
Mike



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#539): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/539__;!!PvDODwlR4mBZyAb0!SmVDQ8ztKqDmWW-d4rCKPGYLpiIJcd0j=
-Uz0tmAQMDopyNK5oMs28ifm1FIs9sSKA6kOkn-JZP3RmGR3$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101198059=
/7559972__;!!PvDODwlR4mBZyAb0!SmVDQ8ztKqDmWW-d4rCKPGYLpiIJcd0j-Uz0tmAQMDopy=
NK5oMs28ifm1FIs9sSKA6kOkn-JZL9J0LZf$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Sep 06 2023 - 11:42:43 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:44:18 MST