Robert,
This is an easy fix. Open the console and find the ACB located in the AQR rack. There is a small opening on the board. Push a pin into the opening and this will press the reset button. This will clear the error. What is happening is that the transmitter is powered off by the ACB (Amplifier control board)
If you have any further issue feel free to contact Bruker center.