AMMRL: TRANS P-DOWN light is on

From: Savage, Veronica <Veronica.Savage_at_bruker.com>
Date: Thu, 3 Sep 2015 19:00:07 +0000

Robert,

This is an easy fix. Open the console and find the ACB located in the AQR rack. There is a small opening on the board. Push a pin into the opening and this will press the reset button. This will clear the error. What is happening is that the transmitter is powered off by the ACB (Amplifier control board)
If you have any further issue feel free to contact Bruker center.

Center_at_bruker.com


Veronica Savage
Center Engineer
Bruker BioSpin Corp. 15 Fortune Drive Billerica, MA 01821
PH#: 978-667-9580 dial # 1 for NMR, then # 2 for SERVICE
EMAIL: center_at_bruker-biospin.com<mailto:center_at_bruker-biospin.com> FAX: 978-667-6168
WEBSITE: http://www.bruker-biospin.com/nmr
Received on Thu Sep 03 2015 - 09:00:22 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 18 2023 - 18:05:09 MST