topspin -r
Will start topspin such that it doesn't point to the last data set.
> On Sep 23, 2014, at 1:41 PM, "Plant,Daniel" <dan_at_mbi.ufl.edu> wrote:
>
> Long ago (uxnmr), there was a user file in curdir after a crash that would hang the system until you removed it...
> Can you delete the file that was up during the outage?
>
> Usually I go back to the nmrsu directory to see if it's just a user problem but this one may hang it up for everyone.
>
>
>
> -----Original Message-----
> From: Robert Honeychuck [mailto:rhoneych_at_gmu.edu]
> Sent: Monday, September 22, 2014 2:05 PM
> To: ammrl_at_ammrl.org
> Subject: AMMRL: TopSpin won't start
>
> Hello all.
>
> A student was in TopSpin on my DPX 300 during a power failure. The console turned itself off. I replaced the battery pack on the 230 V UPS and turned the instrument back on. Now Linux is good, but TopSpin won't start. The Flexlm license seems to be OK. The internal boards are not damaged: fcutest, tcutest, and rcutest find no errors.
>
> The error message mentions a read-only file system, but the permissions look OK in Konqueror.
>
> Here is the error message on attempted startup by user chem. The 1.3 is correct. This is where our topspin resides.
>
> -----
> CPR : Path to prog : "/opt/topspin1.3/prog"
> CPR : Path to exp : "/opt/topspin1.3/exp"
> CPR : Path to conf : "/opt/topspin1.3/conf"
> CPR : waiting for FLEXlm license
> CPR : Your FLEXlm license is valid until 6-nov-2023 CPR : could not correct permissions of current data directory /opt/topspin1.3/prog/curdir/chem: Read-only file system Program is exiting...
> -----
>
> Thank you very much.
>
> Robert Honeychuck
> Dept of Chemistry and Biochemistry
> George Mason U
> Fairfax VA 22030
> rhoneych_at_gmu.edu
>
>