RE: AMMRL: TopSpin won't start

From: Plant,Daniel <dan_at_mbi.ufl.edu>
Date: Tue, 23 Sep 2014 17:50:02 +0000

Hi Robert,
Not sure if I'm on the right track but I think the command to fix it was uxnmr -r (and now probably topspin -r)
Some knowledgable Bruker person will likely be along shortly to straighten us out...

Dan


-----Original Message-----
From: Plant,Daniel [mailto:dan_at_mbi.ufl.edu]
Sent: Tuesday, September 23, 2014 8:12 AM
To: 'Robert Honeychuck'; ammrl_at_ammrl.org
Subject: RE: AMMRL: TopSpin won't start

Long ago (uxnmr), there was a user file in curdir after a crash that would hang the system until you removed it...
Can you delete the file that was up during the outage?

Usually I go back to the nmrsu directory to see if it's just a user problem but this one may hang it up for everyone.



-----Original Message-----
From: Robert Honeychuck [mailto:rhoneych_at_gmu.edu]
Sent: Monday, September 22, 2014 2:05 PM
To: ammrl_at_ammrl.org
Subject: AMMRL: TopSpin won't start

Hello all.

A student was in TopSpin on my DPX 300 during a power failure. The console turned itself off. I replaced the battery pack on the 230 V UPS and turned the instrument back on. Now Linux is good, but TopSpin won't start. The Flexlm license seems to be OK. The internal boards are not damaged: fcutest, tcutest, and rcutest find no errors.

The error message mentions a read-only file system, but the permissions look OK in Konqueror.

Here is the error message on attempted startup by user chem. The 1.3 is correct. This is where our topspin resides.

-----
CPR : Path to prog : "/opt/topspin1.3/prog"
CPR : Path to exp : "/opt/topspin1.3/exp"
CPR : Path to conf : "/opt/topspin1.3/conf"
CPR : waiting for FLEXlm license
CPR : Your FLEXlm license is valid until 6-nov-2023 CPR : could not correct permissions of current data directory /opt/topspin1.3/prog/curdir/chem: Read-only file system Program is exiting...
-----

Thank you very much.

Robert Honeychuck
Dept of Chemistry and Biochemistry
George Mason U
Fairfax VA 22030
rhoneych_at_gmu.edu
Received on Tue Sep 23 2014 - 07:50:09 MST