Hi Simon,

I had exactly that issue a few month ago. In my case it was caused by
poor cooling of the AQS rack due to a couple of failed fans. I changed
all eight fans on the top of the rack, and the issue has gone.

The total cost of repair was about 300 €.

Cheers,

Igor

--
Igor Moudrakovski, PhD
Max-Planck Institute for Solid State Research
Heisenbergstr. 1, 70569 Stuttgart, Germany


On 20/10/2023 16:29, Sham, Shing wrote:
> Hi,
> I have had an intermittent issue from AQS3+, it has happened few times
> in last 2 months.
> 1.  When Topspin tried to wobb or atma(n), it failed.
> 2.  It complained about a power failure in AQS. See pict. #1
> 3.  Then it showed a closed connection between remote host (PC) and
> spectrometer (?). See pict. #2
> 4.  All the components in AQS web service became "unknown". See pict. #3
>
> First time when it happened, I reset the IPSO. It got back to normal,
> Finally, I just clicked on the "confirm" shown in pict.#3 and all the
> components changed back to "OK" instantaneously. It seemed to me it
> meant "Yeah, I know there is a problem on a power supply! I will fix it)
>
> My questions: Is "SLOT 1 RACK 1: Power failed" referred to the PSM-48V
> ac/dc power supply in slot 1?. When it happened, the status LEDs on
> SGUs and others were OK.
>
> Has anyone seen this problem before?
>
> Thanks for your insight in advance.
>
> Simon
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#715): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/715__;!!PvDODwlR4mBZyAb0!U_-HNQRLVEGLiHZ9hHcGlVt3tIefsL7V=
6qozWMbQH_eHohUtDwPg7QmjIWpc0XRU1HUiuFU_Z-3u_3BK863ZYyuRl7zk$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102082269=
/7559972__;!!PvDODwlR4mBZyAb0!U_-HNQRLVEGLiHZ9hHcGlVt3tIefsL7V6qozWMbQH_eHo=
hUtDwPg7QmjIWpc0XRU1HUiuFU_Z-3u_3BK863ZY3lvvgxU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Oct 20 2023 - 08:57:32 MST

This archive was generated by hypermail 2.4.0 : Wed Nov 01 2023 - 14:23:34 MST