[AMMRL] Spellman DF4 HV PS for DNP-NMR Failed to Power On

From: Jerry Hu <jinguihu_at_ucsb.edu>
Date: Fri, 20 Oct 2023 00:23:31 -0700

Dear Colleagues,

The Spellman DF4 high voltage power supply for our DNP-NMR instrument has
stopped working. Basically nothing happens when the CB1 breaker is flipped
up and all the lights and displays on the front panel remain dark. The
Control Power ON button does not light up when pressed. With the top cover
of the unit removed, nothing inside seems to indicate the unit has power.
The Gyrotron control software refuses to initialize because it cannot
detect the high voltage from the unit.

Here is the debugging we have tried:
1. Checked the electricity to the unit, no issue
2. Checked the CB1 breaker, no issue
3. Replaced all the 1W lamps for X-Ray ON, OFF, and Control Power ON on the
front panel
4. Checked the failsafe lamps, no issue
5. Checked all the fuses inside the unit, all good
6. Inspected electronic boards inside, did not see anything obviously wrong.

Your inputs for debugging are appreciated.

Cheers,
Jerry
-------


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#712): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/712__;!!PvDODwlR4mBZyAb0!Tatg31sJEfha5MuHmK1KCNR_FBW4YseP=
QQDipfkvYYUxNxerZt4wJtTOhzQ2E72WF_qDwtqem_W6AZhmkw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102079200=
/7559972__;!!PvDODwlR4mBZyAb0!Tatg31sJEfha5MuHmK1KCNR_FBW4YsePQQDipfkvYYUxN=
xerZt4wJtTOhzQ2E72WF_qDwtqem_WQHLPzlw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Oct 20 2023 - 04:50:07 MST

This archive was generated by hypermail 2.4.0 : Wed Nov 01 2023 - 14:23:34 MST