Re: [AMMRL] updating TopSpin 4.1.4 to TopSpin 4.3.0

From: Bob Berno <bberno_at_mcmaster.ca>
Date: Wed, 18 Oct 2023 13:57:16 +0000

Hello everyone,

The perils of software upgrades!

How many of you have encountered weird error messages and are told to download
and install the latest version of TopSpin? Was this error message due to a
known bug that was fixed in the current release? Maybe, or maybe not.
But applying a software update to a workstation in a multiuser facility is
never as easy and straightforward as they think.

I never bothered with TopSpin 4.2 on any of our spectrometer workstations.
I was waiting for TopSpin 4.2.1, but I guess v4.2 was so bad that they decided
to go straight to v4.3.0 instead. I used v4.2.0 as a datastation and I hated it.
And when I read the release notes for 4.3.0 I was really glad I never bothered
with v.4.2.0 on any of our spectrometer hosts. One of my colleagues described
v4.2.0 as "...the devil, and should never be installed on a spectrometer
workstation."

So far, I have upgraded one spectrometer host to v4.3.0 of TopSpin. It is
our most heavily used spectrometer (in terms of sample throughput), and it
has been working quite well since early August. We did encounter some
"communication" issues at first like have been described, but that was
resolved after a few workstation and console reboots/power cycles. Maybe these
repeated reboots were needed for all of the firmware updates? This spectrometer
workstation runs CentOS 7, as do most of the rest of our workstations.

We have a Fourier 80 benchtop spectrometer running Windows 10 which is also
running TopSpin 4.3.0. It's been online for a few months, and so far it has
been working fine. It was updated to 4.3.0 during the installation, so I can't
comment on how this spectrometer performs compared to earlier versions of TopSpin.

I guess my suggestion is, "if it ain't broke, don't fix it." I've so far had
some good success with v4.3.0 on a spectrometer host, and I plan to start
upgrading the others soon. But if you don't have a pressing need to upgrade,
then don't bother.

Good luck everyone,
Bob.

-----Original Message-----
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Ulrich via groups.io
Sent: October 18, 2023 6:32 AM
To: main_at_ammrl.groups.io
Subject: Re: [AMMRL] updating TopSpin 4.1.4 to TopSpin 4.3.0


Hello All!

I tried TS 4.2 and then 4.3 both are a bit ... a mess. Please stay on 4.1.4
- this one is a bit more stable.

TS 4.3 was so bad - so I had to switch back to 4.2 - but this is not so good.

OK - I use the Linux versions - not the Microsoft Windows 10 or 11 stuff!

Yours sincerly,

Ulrich Haunz

Am Mittwoch, Oktober 18, 2023 01:45 CEST, schrieb "Tom Pratum" <tkp9551_at_gmail.com>:

> Hello All-
> We have recently had a NEO 400 installed with TopSpin 4.1.4 running under
> Windows 10. We are now trying to install 4.3.0, and following the instructions
> one of our faculty members received at a recent training course. We find
> that the new installation is not communicating with the spectrometer, despite
> having all of the correct configuration files from the previous configuration
> in the new configuration directory. We also noted a number of “Windows Defender”
> exceptions when running cf and/or expinstall, so we turned off Windows Defender,
> but still, the new installation is unable to communicate with the spectrometer.
> The old installation still works fine.
>
> Does anyone have any tips on completing this update?
>
> Thanks very much in advance!
>
>
>
> Tom Pratum, Southern Oregon Univ.
>
>
>
>

--
=========================

Ulrich Haunz, Dipl.-Ing. (FH)
NMR-Spektroskopie

Raum L 710, L 807
NMR-Center M520, M521, M527
Postfach M 715
Fachbereich Chemie
Universitaet Konstanz

78457 Konstanz

Tel.: +49 (0)7531 88-4261 -> M521 Bruker Avance NEO 800
Tel.: +49 (0)7531 88-4422 -> L807 JEOL Lambda 400
Tel.: +49 (0)7531 88-4955 -> L710 Bruker AVIII HD 400
Tel.: +49 (0)7531 88-4956 -> M520a Bruker AVIII 600
Tel.: +49 (0)7531 88-4957 -> M527 Bruker AVIII 400 Automation
Tel.: +49 (0)7531 88-4958 -> M520 Bruker AVIII 400 WB
Tel.: +49 (0)170 9153213
FAX : +49 (0)7531 88-3898


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#699): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/699__;!!PvDODwlR4mBZyAb0!VnoZhi0AVJiNWh1G12v2V0hIBdlqUMv_=
xg_zXPXxKksDZj4rWs0XKKSRRorFY-6haQQ7gkWMP2eNfNkX$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102029967=
/7559972__;!!PvDODwlR4mBZyAb0!VnoZhi0AVJiNWh1G12v2V0hIBdlqUMv_xg_zXPXxKksDZ=
j4rWs0XKKSRRorFY-6haQQ7gkWMP_8yMfau$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Wed Oct 18 2023 - 06:57:21 MST

This archive was generated by hypermail 2.4.0 : Wed Nov 01 2023 - 14:23:34 MST