Hi George,
I have opted for ‘Use Lastscale for Calibration’ when in the
integration mode (right click). You have to integrate a relevant spectrum
first and close out of .int mode. Then open a new spectrum, go to .int and
right click for ‘Use Lastscale for Calibration’. I believe
that does exactly what you’re asking.

Good luck,
Vernon

Vernon Russell
vrussell_at_wlgore.com

W. L. Gore & Associates, Inc.
Together, improving life
gore.com http://www.gore.com

From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Sukenick, George
Sent: Tuesday, October 17, 2023 7:02 PM
To: main_at_ammrl.groups.io
Subject: [AMMRL] integration between spectra

I have a researcher who wants to compare integration between different spectra
(all taken with same parameters).

My thought was to integrate without normalizing or referencing the integral,
but it seems that integration (at least in Topspin) does not provide the =
absolute value.

I don’t see anything in the AU program collection either.

Is there a way to obtain this information?

Or while we always use relative integrals – is there a reason not to
use the absolute values?

I suppose that the data could be imported to a spreadsheet and then absolute
intensities summed within a region, but there should be a better way?

Thanks!

    George Sukenick, Ph.D.
       Associate Lab Member/Head
       NMR Analytical Core Facility

       Sloan Kettering Institute
       Memorial Sloan Kettering Cancer Center
       g-sukenick_at_ski.mskcc.org<mailto:g-sukenick_at_ski.mskcc.org>

==========================
==========================
===================

https://www.gore.com/sites/g/files/ypyipe116/files/2022-01/GORE-Logo-Email-Signature_0.png
For information about our privacy practices, see our Privacy Notice
https://www.gore.com/privacy

This email may contain trade secrets or privileged, undisclosed or otherwise
confidential information. If you have received this email in error, you are
hereby notified that any review, copying or distribution of it is strictly
prohibited. Please inform us immediately and destroy the original transmittal.
Thank you for your cooperation.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#698): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/698__;!!PvDODwlR4mBZyAb0!Ve21FiHkgvtqv-lceR-kknLO-fVFd76J=
c0iqmd273hdHVQUjjJN4xHLBnI-dYkwsKHIpyI9EG6hvlenw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102029964=
/7559972__;!!PvDODwlR4mBZyAb0!Ve21FiHkgvtqv-lceR-kknLO-fVFd76Jc0iqmd273hdHV=
QUjjJN4xHLBnI-dYkwsKHIpyI9EG19OGXnd$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Oct 18 2023 - 05:05:15 MST

This archive was generated by hypermail 2.4.0 : Wed Nov 01 2023 - 14:23:34 MST