FW: AMMRL: question about 1H/19F switch

From: Gabriel Gonzàlez <ggonzalez_at_ICIQ.ES>
Date: Tue, 23 Oct 2018 08:36:16 +0000

Dear all,

First of all, thank you very much to all the people who have provided an answer to our query.

The summary is very easy, the unanimous opinion is that is not possible to decouple 1H and 19F using a 1H/19F switch. To do that an additional channel with and additional 19F amp would be needed.

Thanks again to all the people that showed an interest in this subject.

Best wishes,

Gabriel


[Descripción: logoICIQ]

Dr. Gabriel González Ferrer
Responsable RMN / NMR Manager
Institut Català d'Investigació Química (ICIQ) /
Institute of Chemical Research of Catalonia (ICIQ)
The Barcelona Institute of Science and Technology
Av. Països Catalans 16 - 43007 Tarragona (Spain)
Phone +34 977920207 / +34 686669420 - Fax +34 977920224
ggonzalez_at_iciq.es<mailto:ggonzalez_at_iciq.es> - www.iciq.es<http://www.iciq.es/>








This e-mail contains proprietary information whose content can be protected by law. If you are not the intended recipient, you may not use, disclose, distribute, copy or print this e-mail. This confidential, personal and not transferable e-mail is intended for the addressee only and is legally protected. Any disclosure, unauthorized use or forwarding, in whole or in part is not permitted. If you are not the intended recipient of this e-mail message and its contents, please notify the sender immediately and delete this message and all its attachments subsequently without reading or record, in whole or in part. Thank you.

Aquest correu electrònic conté una informació propietària el contingut del qual pot estar legalment protegit. Si vostè no és el destinatari previst, no pot usar, divulgar, distribuir, copiar o imprimir aquest correu electrònic. Aquest missatge electrònic està dirigit únicament a la/les direcció/ns indicada/es anteriorment; el caràcter confidencial, personal i intransferible del mateix està protegit legalment. Qualsevol revelació, ús o reenviament no autoritzat, complet o en part, està prohibit. Si ha rebut aquest missatge per equivocació, notifiqui-ho immediatament a la persona que ho ha enviat i esborri el missatge original juntament amb els seus fitxers annexos sense llegir-lo ni gravar-lo, totalment o parcialment. Gràcies.

Este correo electrónico contiene una información propietaria cuyo contenido puede estar legalmente protegido. Si usted no es el destinatario previsto, no puede usar, divulgar, distribuir, copiar o imprimir este correo electrónico. Este mensaje electrónico está dirigido únicamente a la/s dirección/es indicada/s anteriormente; el carácter confidencial, personal e intransferible del mismo está protegido legalmente. Cualquier revelación, uso o reenvío no autorizado, completo o en parte, está prohibido. Si ha recibido este mensaje por equivocación, notifíquelo inmediatamente a la persona que lo ha enviado y borre el mensaje original junto con sus ficheros anexos sin leerlo ni grabarlo, total o parcialmente. Gracias.



From: Gabriel Gonzàlez [mailto:ggonzalez_at_ICIQ.ES]
Sent: Thursday, October 18, 2018 5:57 AM
To: ammrl_at_ammrl.org<mailto:ammrl_at_ammrl.org>
Subject: AMMRL: question about 1H/19F switch

Dear all,

We have a Bruker instrument with a 1H/19F/Broadband 3 channel probe.

As 1H and 19F go through the same preamplifier (proton) they are controlled by a switch which alternates one nucleus or the other, allowing us to do 1H{19F} or 19F{1H}.

The question is, has somebody ever tried to use this kind of 1H/19F switch to decouple 1H and 19F consecutively? Would that be possible?

The idea would be to acquire 13C spectra with 1H and 19F decoupling without having to purchase another channel/preamplifier.

Thanks in advance for your answers,



[Descripción: logoICIQ]

Dr. Gabriel González Ferrer
Responsable RMN / NMR Manager
Institut Català d'Investigació Química (ICIQ) /
Institute of Chemical Research of Catalonia (ICIQ)
The Barcelona Institute of Science and Technology
Av. Països Catalans 16 - 43007 Tarragona (Spain)
Phone +34 977920207 / +34 686669420 - Fax +34 977920224
ggonzalez_at_iciq.es<mailto:ggonzalez_at_iciq.es> - www.iciq.es<http://www.iciq.es/>








This e-mail contains proprietary information whose content can be protected by law. If you are not the intended recipient, you may not use, disclose, distribute, copy or print this e-mail. This confidential, personal and not transferable e-mail is intended for the addressee only and is legally protected. Any disclosure, unauthorized use or forwarding, in whole or in part is not permitted. If you are not the intended recipient of this e-mail message and its contents, please notify the sender immediately and delete this message and all its attachments subsequently without reading or record, in whole or in part. Thank you.

Aquest correu electrònic conté una informació propietària el contingut del qual pot estar legalment protegit. Si vostè no és el destinatari previst, no pot usar, divulgar, distribuir, copiar o imprimir aquest correu electrònic. Aquest missatge electrònic està dirigit únicament a la/les direcció/ns indicada/es anteriorment; el caràcter confidencial, personal i intransferible del mateix està protegit legalment. Qualsevol revelació, ús o reenviament no autoritzat, complet o en part, està prohibit. Si ha rebut aquest missatge per equivocació, notifiqui-ho immediatament a la persona que ho ha enviat i esborri el missatge original juntament amb els seus fitxers annexos sense llegir-lo ni gravar-lo, totalment o parcialment. Gràcies.

Este correo electrónico contiene una información propietaria cuyo contenido puede estar legalmente protegido. Si usted no es el destinatario previsto, no puede usar, divulgar, distribuir, copiar o imprimir este correo electrónico. Este mensaje electrónico está dirigido únicamente a la/s dirección/es indicada/s anteriormente; el carácter confidencial, personal e intransferible del mismo está protegido legalmente. Cualquier revelación, uso o reenvío no autorizado, completo o en parte, está prohibido. Si ha recibido este mensaje por equivocación, notifíquelo inmediatamente a la persona que lo ha enviado y borre el mensaje original junto con sus ficheros anexos sin leerlo ni grabarlo, total o parcialmente. Gracias.



image001.jpg
(image/jpeg attachment: image001.jpg)

Received on Mon Oct 22 2018 - 22:36:33 MST

This archive was generated by hypermail 2.4.0 : Tue Jun 20 2023 - 16:35:20 MST