Hi Holger,
have you tried the ICONNMR web interface. You first need to activate
this mode in the configuration window.
The users then can submit their samples from a standard web browser. The
only requirement is that on the spectrometer the automation window has
been opened, the run does not need to be started. Once you want to
switch to automation mode you simply start the run.
There is even a modified version for data entry from a mobile device.
Check out pages 69 ff in the ICONNMR setup manual.
Regards
Clemens
On 10/1/2018 5:04 PM, Frank H Foersterling wrote:
> Everybody,
>
> In order to avoid reinventing the wheel I figure it is best to place the
> question here. We operate our new Avance500 spectrometer (Topspin
> 3.5pl7 with Sample CASE) currently with individual users logging into
> their accounts and running spectra manually or through ICONNMR. We have
> ICONNMR set up such that any user can log into an existing queue and
> enter and submit samples. I have defined different groups of users with
> access to different sets of experiments and different permissions as to
> which parameters they are allowed to modify.
>
> What we would like to do is to allow users to submit samples and enter
> them into a queue during the day for overnight/weekend runs while the
> spectrometer is used in manual mode, i.e. without ICONNMR running on the
> spectrometer. I am aware there is the option to import existing queues
> from ICONNMR or from spreadsheets, so the question is what is the most
> straightforward way to set up a way for people to enter experiments to a
> spreadsheet or ICONNMR queue-file at an external computer or through a
> web interface while using the information defined in the
> $TOPSPIN/conf/instr/spec/inmrusers directory on the spectrometer,
> ideally using existing software. like a second copy of topspin running
> on a separate workstation.
>
>
> Thank you very much for any advise.
>
> Holger Foersterling
>
> University of Wisconsin-Milwaukee
>
>
--
==============================================================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive Phone:978 667 9580 ext 5144
Billerica MA 01821 skype_ID: canklin
web: www.bruker.com e-mail: clemens.anklin_at_bruker.com
==============================================================
Received on Tue Oct 02 2018 - 04:36:58 MST