Dear all,
Our IT expert fixed the problem of the flexlm server, and it worked now.
I'd like to thank Clemens and Allen from Bruker, and all of you who
responded: Stuat, Richard, Dan, Jiejun, Bob, George, Weiguo, Stefano,
Dejian, Leslie, Chris.
Thank you very much!
Jinghua
>Dear all,
>
>The Topspin1.3pl6 on our Avance500 quit by itself and wouldn't restart
>from then. I have included the error message here. Can anyone help?
>
>
>Thank you in advance!
>
>Jinghua Yu
>
>The Sanford Burnham Institute
>858-646-3100, ext. 3630
>
>
>
>
>the error message:
>
>
>CPR : checking connection to spectrometer ( may last a minute )
>OrbManager: Error: CORBA::SystemException: INITIALIZE
>(IDL:omg.org/CORBA/INITIAL IZE:1.0)
>
>Warning: an important component (ORB) failed to initalise properly.
>You may want to inspect the logfile at
>'/opt/topspin/prog/curdir/ucsd/history.tr affic'.
>
>Program is exiting...
>
>Hit ENTER to continue ...
>
>
>And in /opt/topspin/prog/curdir/ucsd/history.traffic
>
>---------------------------------------------------
>History of CORBA traffic CPR --> GUI front-ends
>----------------------------------------------------
>Debug level is 0
>
>CORBA error: initialization of ORB failed. No client communication
>possible.
>
>
>Thank you!
>
>Jinghua Yu
>
>The Sanford Burnham Institute
>858-646-3100, ext. 3630
>
>
Received on Tue Oct 11 2011 - 07:48:28 MST