This is a known problem with this MS patch. See the following:
http://news.zdnet.co.uk/software/windows/0,39020396,39230991,00.htm
Gary Strahan
USDA, ARS, NAA, ERRC
Dr. Matthias Findeisen wrote:
> I installed the WindowsXP safety update KB 902400 (MS05-051) on a PC,
> running TopSpin (1.3) (german Version of XP)
>
> Afterwards the spectrometer didn't work.
>
> The reason for: the windows firewall blocked the nfs-server on the
> second network card (to the spectrometer).
> The MIPS console RS-232-monitor told: "nfs server not responding"
>
> The Bruker script :"setfirewall.cmd -auto" passed wthout error, the
> firewall exception entries were correct.
> But this didn't solve the problem.
>
> Workaround:
> de-aktivate the firewall on the spectrometers network card.
> Or: de-install KB 902400.
>
>
> Matthias Findeisen
> Univ. Leipzig
> Group of Prof. Berger
>
Received on Thu Oct 20 2005 - 18:25:20 MST