Acquistion error, code = 900

From: Jack Howarth <howarth_at_nitro.med.uc.edu>
Date: Mon, 18 Oct 2004 10:12:47 -0400 (EDT)

   On one of our Sun Blade 1500's running VNMR J 1.1D we have been
seeing 2D and 3D experiments fail with an error message of the form...

Sun Oct 17 20:08:46 2004: FID 6641: Acquistion error, code = 900

Varian has this listed in their VNMR J 1.1 bug list as an unresolved
bug. However I am told that the application lab at Varian hasn't seen
this error message in a long time. Interestingly I haven't seen this
problem on a second console running VNMR J 1.1D. The only clear difference
I can find is that the console that is immune to the problem has been
used with dsp='r' and the other has dsp='n'. I have recommended that the
users try dsp='r' on the problem console and see if that resolves it.
Has anyone else seen these 'code = 900' errors recently? If I understand
the error code it implies that an undefined error has occured so it
may be hard to debug.
                Jack
------------------------------------------------------------------------------
Jack W. Howarth, Ph.D. 231 Albert Sabin Way
NMR Facility Director Cincinnati, Ohio 45267-0524
Dept. of Molecular Genetics phone: (513) 558-4420
Univ. of Cincinnati College of Medicine fax: (513) 558-8474
Received on Mon Oct 18 2004 - 09:34:27 MST

This archive was generated by hypermail 2.4.0 : Thu Jun 08 2023 - 19:19:38 MST