Varian Setup Macro and Poltergeists

From: Neil Jacobsen <neil_at_u.arizona.edu>
Date: Tue, 31 Oct 2000 11:49:19 -0700

We have had problems with users of our Varian Unity-300 who lock and shim,
getting a nice high lock signal, and then start an experiment with "go" and
see the lock plummet. Sometimes this happen even within the lock and shim
menu, when switching from "lock" to "shim" and back. The problem seems to
be due to changing values of lock power and lock gain, and possibly lock
phase, rather than any change in shims.
        In the process of searching for the poltergeist, I looked at the "setup"
macro in the maclib and wondered if this is executed when you enter "su" or
"go", or is it executed only when you select "Main Menu" and "Setup" with
the Menu Buttons. Then I got the idea that we could write a "Bruker-like"
macro, similar to the "edlock" feature in XWinNMR, which would load
appropriate values of lock power, Z0, lock gain and lock phase for a given
solvent. The problem is that a Z0 value, being essentially a shim, is not
used until you enter "su", which then sets a lot of other things I don't
want changed.
        Could someone explain this to me? Is such a lock solvent macro possible
in VNMR?