> I was asked an interesting question by an Organic Chemistry faculty
> member. In a chiral molecule an ethyl group, CH3-CH2, can have two
> different chemical shifts for the two geminal protons on the CH2
> group. But I've never seen a double-doublet for the methyl group,
> always a triplet. Why can't the two vicinal J values be different
> if the two protons on the CH2 are not equivalent?
>
It's because of the very fast rotation of the methyl group. This
rotation is so fast that under the usual experimental conditions, the
three protons experience an averaged vicinal couplings with both of
the two diastereotopic protons, and the averaged 3J are usually the
same. Hope my answer helps you and your colleague.