I have succeeded to make xwinnmr 3.5 on centos 6.5. You should succeed to
do the same with topspin 1.3
Le Tue Nov 25 2014 at 18:44:34, Surface, James A <surface_at_ppg.com> a écrit :
> All,
>
> I'm writing to see if anyone has successfully installed and controlled a
> spectrometer using Topspin 1.3 (pl 10) on a Windows 7/8 machine using
> compatibility mode, Virtual Machine, or other method. I've got a legacy
> console (Bruker Avance) that, even with some board upgrades, will only
> support up to TopSpin 1.3 (per Bruker). Currently its running XWin-NMR on a
> Silicon Graphics O2 computer. We need to upgrade the computer (its getting
> harder to find replacement parts for it) and in doing so will need to
> upgrade TopSpin (XWin-NMR). IT would prefer a newer machine (no surprises
> there) and Bruker says the only Windows-based OS they can install TopSpin
> 1.3 on is Windows XP, which is not getting security updates anymore. Its
> also harder to justify spending money on a machine that is that deprecated.
>
> Thanks in advance for any suggestions you might have!
>
> Sincerely,
>
> Andy Surface
> NMR Lab Manager, Monroeville Chemical Center, PPG Industries
>
> ________________________________
>
> CONFIDENTIALITY NOTICE:
> This email (including any attachments) is intended for the sole use of the
> intended recipient/s and may contain confidential information, which also
> may be legally privileged. Any reliance upon, access to, review,
> disclosure, copying, forwarding or other distribution of any or all of the
> contents in this message by others who are not the intended recipients is
> STRICTLY PROHIBITED. If you are not the intended recipient, please delete
> the message and all copies and confirm to the sender by email. Your
> cooperation is appreciated.
>
>
Received on Wed Nov 26 2014 - 10:19:16 MST