AMMRL: Safe power levels for CW 1H on DRX-500/600

From: Neil Jacobsen <neil_at_email.arizona.edu>
Date: Tue, 18 Nov 2008 13:43:58 -0700

I have some users trying to evaluate MRI contrast agents and they use
a technique of very long (4 s) cw irradiation at relatively high
power (30 dB) in a series of 1D experiments. My rule has always been
to limit presaturation (1.5 s. irradiation) to 56 dB or 50 dB in
extreme cases. I also tell people to monitor the lock level and the
probe (thermocouple) temperature as soon as the experiment is started.

These seem like dangerous levels for such high duty cycles, but I
don't like to get in the way of research. I'd like to know what
people think in terms of possible damage to probe and
amplifiers. The user says he has done this routinely on "other
machines". The probes we are using are Nalorac (Varian) 3-res 3-axis
gradient HCN and Bruker Dual (non grad) CH, all 5 mm.

Thanks.

Neil