Re: AC power requirements: Varian

Alan Olson (awo@helix.nih.gov)
Wed, 12 May 1999 07:38:00 -0400

Bill -

Yes, I was also surprised at the 220 vac requirement. No, the console
won't be happy at 208 vac. I called up the local electrical supply company
and they had a small read-to-go autotransformer to fill the bill, about 6
inches on a side for about $150, and it works fine. Try your local
supplier, and call if you need more details.

- Alan Olson

At 1:52 PM -0400 5/11/99, Bill Gurley wrote:
>We have just purchased two new Varian instruments, which have
>not yet been installed. During the pre-installation site visit, we
>were surprised to learn that our standard 208V single-phase power
>may not be adequate. It seems that Varian's console design
>requires 220Vac, plus-or-minus 7%, which puts 208V near the
>edge of what is acceptable.
>
>I would like to ask AMMRL list members who may have purchased
>Varian instruments in recent years how they may have dealt with
>this issue, or perhaps if you just ignored it and are operating
>happily with 205V or 208V! It seems that we have these choices:
>
>1. Do nothing and take our chances with the lower voltage.
>2. Purchase step-up transformers to give us a true 220V. (Varian
>sells these for about $1000 each. Anyone know of other sources?)
>3. Purchase a UPS system, which would provide the required
>220V output, along with the obvious other advantages.
>
>Obviously, the third choice is the best one, but at an expensive
>cost. We have pretty much exhausted our available funds just to
>make this purchase, so we're trying to avoid choice number 3.
>
>Your input will be appreciated.
>
>------------------------------------------------------------------------
> Bill Gurley, Supervisor of Technical Services bgurley@utk.edu
>------------------------------------------------------------------------
> Department of Chemistry, University of Tennessee
> Knoxville, TN 37996-1600 phone 423-974-3145
>------------------------------------------------------------------------

------------------------------------------------