Besides using unique shim matrix files and H-1 or C-13 job parameter files
for different solvents to allow the students the easiest start possible I
found it to be the simplest solution to have every student, after he/she
has finished his/her spectrometer session, execute a little EXE program
("EXE FINISH"), which I wrote for our Bruker AM and ACs.
It resets the spectrometer to a predefined condition:
- set spinning speed and turn on sample spinning
- turn off decoupler
- load proton parameters
- load shim file for the standard sample
- set the Bo shim and lock the sample
- start auto shim procedure
- display a message about our user policies
and other neat (probably useless) stuff
The whole procedure is started with one command and the spectrometer will
be available for the next user after 1-2 minutes.
---------------------------------------------------------------------